09931528A
16
List
Function
Customize
Selecting this means that whenever you click on a well to set a
measurement point, the measurement point is set exactly where you
clicked, rather than being automatically centered on the well.
Normally the customize option is used for a large well (eg. 4 or 16)
Calculation Mode
Select Average or Sum.
Average: If selected, multiple reads for a given well are averaged.
Sum: If selected, multiple reads for a given well are summed
(integrated). The results spreadsheet is updated when all readings
are performed for a given well.
Cycle Number
This is the number of measurement cycle.
Time Interval
The time interval is the time between last measurement of the
current cycle and the first measurement of next cycle.
Before the next cycle measurement, a message window will pop up
as shown below. Click Skip to measure immediately.
5.
After setting the parameters in the Data Collection, select Sample Tables tab. Add the sample
numbers and select the each sample type.
Summary of Contents for FL 6500