Never use too much water.
•
Do not bend the cuff or sleeve. Wrap the inside to the outside.
•
Do not dismantle the unit, remove the cuff or try to repair yourself. Contact your distributor.
•
Do not operate the unit under severe climatic circumstances or at extreme temperatures,
humidity or direct sunlight.
18. Troubleshooting and Error Messages
Trouble/Error Possible
Cause
Solution
No indication when powered
on
•
Batteries are not properly
installed
•
Drained batteries
•
Check the polarity of the
batteries
•
Remove the batteries and
wait 10 seconds before
reinstalling
Err 0
•
No pulse or detected pulse
too low
•
Refit the cuff and repeat
the measurement
Err 1
•
Cuff is leaking air or
inflation is too slow
•
Replace the batteries or
contact your distributor
Err 2
•
Cannot measure systolic
•
Refit the cuff and repeat
the measurement
Err 3
•
Cuff pressure is too high
•
Refit the cuff and repeat
the measurement
Err 4
•
Cannot measure systolic
or is outside of range
•
Refit the cuff and repeat
the measurement
Err 5
•
Deflation is too fast
•
Contact your distributor
Err 6
•
Deflation is too fast
•
Contact your distributor
Err 7
•
Systolic is outside of range
•
Refit the cuff and repeat
the measurement
Err 8
•
Main pressure is outside of
range
•
Refit the cuff and repeat
the measurement