PerfectSpin 24 R Service Manual
Relay & Capacitor Modules Disassembly
4.10.
33
1. After removing the top case, disconnect
the cables as shown.
2. Unscrew the screws and disconnect the
cables as shown above.
3. Remove the screw as shown.
4. Use tool to pull out the Compressor
Starting Relay as shown above.
5. Remove the two screws as shown above.
6. Relay Capacitor module is disassembled.
Summary of Contents for PerfectSpin 24 R
Page 1: ...Refrigerated Microcentrifuge PerfectSpin 24 R Service Manual Ver 1 0...
Page 3: ...PerfectSpin 24 R Service Manual ii...
Page 8: ...PerfectSpin 24 R Service Manual Figure 3 Wire Connection Drawing II 5...
Page 10: ...PerfectSpin 24 R Service Manual Figure 4 Connector Locations of Power Board PCB 7...
Page 26: ...PerfectSpin 24 R Service Manual 23 11 Lid cover is fully disassembled as shown above...
Page 39: ...5 Spare Parts List 5 Spare Parts List Lid CoverAssembly 5 1 36...
Page 41: ...5 Spare Parts List Top Case Assembly 5 2 38...
Page 43: ...5 Spare Parts List Base Assembly 5 3 40...
Page 46: ...43 PerfectSpin 24 R Service Manual 5 4 Solenoid Module Assembly...
Page 48: ...PerfectSpin 24 R Service Manual Motor ModuleAssembly 5 5 45...
Page 50: ...PerfectSpin 24 R Service Manual 5 6 Final Assembly 47...
Page 52: ...PerfectSpin 24 R Service Manual Wire Drawing 5 7 49...