PerfectSpin 24 R Service Manual
The coolant used in PerfectSpin 24 R is R134a which is an environment-friendly
coolant
not deplete ozone. The filling amount is 86 gm.
and does
Figure 6. Refrigeration System Scheme
3. Troubleshooting
Please refer the following problem and action list to diagnose and resolve the problems.
More troubleshooting flows are described in Section 3.2.
3.1.
Problem List
9
Problem
Cause
Action
The display is off even
when power is
switched on.
1. Power is not reaching the
system.
2. Power
cord
is
not
plugged into the socket
properly.
3. Power fuse is blown out.
1. Check
power
source
voltage.
2. Reconnect
the
power
cord.
3. Replace the power fuse.
4. Return
the
unit
for
service.
Summary of Contents for PerfectSpin 24 R
Page 1: ...Refrigerated Microcentrifuge PerfectSpin 24 R Service Manual Ver 1 0...
Page 3: ...PerfectSpin 24 R Service Manual ii...
Page 8: ...PerfectSpin 24 R Service Manual Figure 3 Wire Connection Drawing II 5...
Page 10: ...PerfectSpin 24 R Service Manual Figure 4 Connector Locations of Power Board PCB 7...
Page 26: ...PerfectSpin 24 R Service Manual 23 11 Lid cover is fully disassembled as shown above...
Page 39: ...5 Spare Parts List 5 Spare Parts List Lid CoverAssembly 5 1 36...
Page 41: ...5 Spare Parts List Top Case Assembly 5 2 38...
Page 43: ...5 Spare Parts List Base Assembly 5 3 40...
Page 46: ...43 PerfectSpin 24 R Service Manual 5 4 Solenoid Module Assembly...
Page 48: ...PerfectSpin 24 R Service Manual Motor ModuleAssembly 5 5 45...
Page 50: ...PerfectSpin 24 R Service Manual 5 6 Final Assembly 47...
Page 52: ...PerfectSpin 24 R Service Manual Wire Drawing 5 7 49...