3. Troubleshooting
Error Messages
3.2.
12
Message
Cause
Action
“
” Error.
1.
Press
(START/STOP) knob
when Lid is open.
2.
Defective lid lock sensor.
1.
Close the lid.
2.
Change
Lid
lock
module.
3.
Return the unit for
service.
“
” Error.
(rotor
imbalance)
1.
Tubes
are
not
inserted
symmetrically in rotor holes.
2.
Sample liquid in tubes not properly
balanced.
3.
Defective or improperly adjusted
balance sensor.
4.
Rotor is stuck when a run starts.
1.
Load
tube
symmetrically.
2.
Make sure that the
same volume of liquid
is in each tube.
3.
Adjust balance sensor.
4.
Change Rotor or Motor.
5.
Return the unit for
service.
Summary of Contents for PerfectSpin 24 R
Page 1: ...Refrigerated Microcentrifuge PerfectSpin 24 R Service Manual Ver 1 0...
Page 3: ...PerfectSpin 24 R Service Manual ii...
Page 8: ...PerfectSpin 24 R Service Manual Figure 3 Wire Connection Drawing II 5...
Page 10: ...PerfectSpin 24 R Service Manual Figure 4 Connector Locations of Power Board PCB 7...
Page 26: ...PerfectSpin 24 R Service Manual 23 11 Lid cover is fully disassembled as shown above...
Page 39: ...5 Spare Parts List 5 Spare Parts List Lid CoverAssembly 5 1 36...
Page 41: ...5 Spare Parts List Top Case Assembly 5 2 38...
Page 43: ...5 Spare Parts List Base Assembly 5 3 40...
Page 46: ...43 PerfectSpin 24 R Service Manual 5 4 Solenoid Module Assembly...
Page 48: ...PerfectSpin 24 R Service Manual Motor ModuleAssembly 5 5 45...
Page 50: ...PerfectSpin 24 R Service Manual 5 6 Final Assembly 47...
Page 52: ...PerfectSpin 24 R Service Manual Wire Drawing 5 7 49...