Operation
2
0
1
7
-0
5
61
6.2
Measured Value
The
Measured Value
function displays the current value and status of the variables selected in
the
Device Variable Mapping
dialog.
The following information is displayed:
■
Primary Variable
■
Secondary Variable
■
Tertiary Variable
■
Quaternary Variable
■
Loop-Current
5 s ... 10 s
Join wireless
network
■
Keep button pressed for 5 seconds.
■
After 5 sesconds the yellow LED starts flashing at 3 Hz.
■
Release button.
■
The yellow LED extinguishes.
■
The adapter starts joining.
10 s ... 15 s Start
configuration
mode
■
Keep button pressed for 10 seconds. After 5 seconds the
yellow LED starts flashing at 3 Hz. After 10 seconds the red
LED starts flashing at 3 Hz.
■
The adapter enters configuration mode.
■
Release button.
■
Configuration mode is retained until there was no HART
communication for 300 seconds (configurable) or until
pushbutton is pressed again for 10 seconds ... 15 seconds.
■
The red LED extinguishes.
> 15 s
No function
Inserting the battery while the button is already pressed
The following functions are available when connecting the battery while the button is already
pressed. The duration starts from the moment the battery is connected (red LED lights).
< 10 s
No function
■
The red LED lights until the pushbutton is released.
10 s ... 15 s Configuration
reset to
default
settings
■
Keep button pressed for 10 seconds. The red LED lights
during this time.
■
After 10 seconds the red LED starts flashing at 2.5 Hz.
■
Release button to start the reset.
20 s ... 25 s Firmware and
configuration
reset to
default
settings
■
Keep button pressed for 20 seconds. During this time the red
LED lights for 10 seconds. Then the red LED starts flashing at
2.5 Hz. After this, the red LED lights again.
■
After 20 seconds the red LED starts flashing at 1.25 Hz.
■
Release button to start the reset.
> 25 s
No function
■
The red LED remains lit until the pushbutton is released
Duration
Function
Procedure