SPECIFICATIONS
8
2
Specifications
Item
Specification
F
or
USA & C
anada
E
urope
& O
ceania
Power Requirements
Voltage
120 VAC (NTSC MODEL)
230 V (PAL MODEL)
Frequency
50 – 60 Hz
Power consumption
1.5 A
0.8 A
Voltage fluctuation
+/– 10%
Operating Environment
Temperature
Relative humidity
Air pressure
10 ~ 40 °C
30 ~ 85 %
700 – 1060 hPa
Storage/Transport Environment
Temperature
Temperature
Relative humidity
Air pressure
–20 ~ 60 °C
0 ~ 85 %
700 ~ 1060 hPa
Illumination
Lamp
Metal Halide Lamp
MSCR 70-50E
Lamps average life span
200 hours, continuous use
Colour temperature
12,000 K
Lighting format
Switching regulator with continuous illumination
Brightness control
Selection – Automatic or Manual
Automatic iris
Servo type
Scope Compatibility
Pentax colour video endoscopes
NTSC M
odel
PAL M
odel
Pentax fiberscopes
With use of appropriate fiberscope video adapter module
NTSC Modell
PAL Modell
Other manufacturers’ fiberscopes
With use of appropriate fiberscope video adapter module and
appropriate eyepiece/light guide adapters
Air Feed System
Air pump system
Electro-magnetic vibrator system
Pressure setting at zero flow rate
0.38 ~ 0.42 kg/cm2
(37 ~ 41 kPa)
0.42 ~ 0.63 kg/cm2
(41 ~ 62 kPa)
Standard air feed volume at inlet of water
bottle
3.1 ~ 5.3 L/min
4.5 ~ 8.0 L/min
Water Feed System
Water bottle assembly pressurized by pump
Bottle capacity = 250 mL
Brightness Control System
Automatic
Manual
Selection - Average or Centre
+/– 5 step adjustment
Colour System
Colour correction
Red +/– 5 steps
Blue +/– 5 steps
Freeze Function
Live video image provided when freeze
mode activated
Cooling
Forced air cooling
Video Outputs
2 sets: RGBS (NTSC or PAL), 9-pin D-Sub female connectors
2 sets: Composite (NTSC or PAL), BNC connector
2 sets: Separate Video (Y/C), 4-pin female connector
1 set: Computer, 9-pin D-Sub female connector
Classification as Electro-Medical
Equipment
Type of protection against electric shock
Class I equipment, three-pin plug
Degree of protection against electric shock
BF Type (Body Floating), using insulated endoscope.
Use on heart is prohibited
Degree of explosion proofing
Do not use in potentially flammable surroundings
Audible Noise
Sound pressure level
< 70 dB (A) (Based on ISO 7779)
Electro-Magnetic Compatibility
CLSPR1 1
EN 60601-1-2
Compliance
Designed in accordance with
UL 2601-1
EN 60601-1
Size
Dimensions
Width = 380 mm Height = 153 mm Depth = 405 mm
Weight
Main Body = 13.5 kg (29.8 lbs)