Cascade Comfort System  

 

1.

 

Start the air blower by pressing the Aqua-Aire ON/OFF button. 

WARNING

 

In the following bathing instructions it is recommended that the resident be rinsed with 
the shower sprayer. Before doing so each time, make sure the temperature is between 
95 and 105 degrees Fahrenheit (35 to 41 degrees Celsius) by viewing the temperature 
readout. As a further precaution, always test the water on the inside of your wrist 
before turning the shower sprayer toward the resident. Failure to follow the 
precautions in this warning could result in a burn injury to the operator or the resident.  

2.

 

Bathe the resident’s upper body as follows: 

a)

 

Begin gently washing the resident’s upper body. Pay special attention to folds or creases in 
the skin and under the arms.  

3.

 

Rinse the upper body with the shower sprayer as follows. 

a)

 

Pull out the Hand Sprayer. 

b)

 

Press the “HAND SPRAYER ON/AUTO OFF” button. This starts the water flow to the hand 
sprayer for a set period of time. Then the water flow is automatically stopped. It can be 
restarted at any time by pressing the “HAND SPRAYER ON/AUTO OFF” button. 

Note.

 This time period is adjustable, up to 10 minutes, by the maintenance department. 

c)

 

Direct the sprayer into the spa, 

not at the resident. 

Press the trigger on the hand sprayer and 

adjust the mixing valve, if needed, to the normal bathing range of 95 to 105 degrees 
Fahrenheit (35 to 41 degrees Celsius). Monitor the temperature via the Fill/Shower 
Temperature readout. As a further precaution, always test the water on the inside of your wrist 
before turning the sprayer toward the resident. 

d)

 

The Aqua-Aire action in the Cascade Comfort will help clean the resident’s.  

Transferring Out Of Cascade Comfort 

After the bath has been given, you are now ready to transfer the resident out of the spa. 

1.

 

Drain the spa

 by pulling the drain plug from the drain. To avoid water overflow start draining 

the spa before bringing the spa to the full upright position.  

 

WARNING 

Failure to have the spa completely drained before unlocking the door could cause 
injury to the operator and resident. 

2.

 

Press and hold the down button until the spa is in the full upright position. 

3.

 

Unlock door by pulling up on the door handle release and pulling up on the lever at the same 
time. 

4.

 

Open door to widest position. 

5.

 

Dry off resident. 

6.

 

Carefully maneuver resident’s legs around and use proper nursing techniques to transfer to a 
wheelchair. 

7.

 

If using a sit-to-stand lift or total lift, follow the manufacturer’s instructions on the use of that 
particular lift. 

8.

 

You are now ready to disinfect the Cascade Comfort Spa. 

Summary of Contents for Cascade Comfort 560010-X

Page 1: ...e Operation Daily Maintenance Instructions PENNER PATIENT CARE INC Box 523 101 Grant St Aurora NE 68818 560750 Revision A 12 11 15 1 866 PENNERS 1 866 736 6377 1 800 732 0717 E mail Address pennersale...

Page 2: ...COMFORT SPA CONTROLS 5 CASCADE DOOR 6 LOCKING CASCADE DOOR 6 UNLOCKING CASCADE DOOR 6 SYSTEM PREPARATION BEFORE BATH 7 USING CASCADE COMFORT BATHING SYSTEM 8 BATHING PROCEDURE 8 TRANSFERRING OUT OF C...

Page 3: ...tion information Failure to obey a caution warning may result in damage to the equipment and may void the warranty Left or Right When the terms left or right are used with reference to the spa this me...

Page 4: ...tient weight rating 400 lbs Maximum Upright position to reclined position angle approximately 30 Reclining actuator rating 1 minute ON 9 minutes OFF All temperature read out values are accurate to 2 F...

Page 5: ...ECT JETS 3 HAND SPRAYER ON AUTO OFF 4 FILL SHOWER TEMPERATURE DISPLAY 5 FILL SHOWER TEMPERATURE CONTROL MIXING VALVE 6 SPA TEMPERATURE DISPLAY 7 AQUA AIRE ON OFF 8 TUB FILL ON OFF 9 HAND SHOWER 10 SPO...

Page 6: ...cade Door Figure 2 1 With door handle in the upright position close the door tight against the spa 2 When the door is against the spa push the door handle to the left or down so the door handle is in...

Page 7: ...uch as a washcloth towels and liquids such as shampoo and body wash You are now ready to bring the Resident or Patient to the Bathing Area with your Patient Transfer System or Wheel Chair 2 At the bat...

Page 8: ...st the water temperature by turning the Temperature Control knob to normal bathing temperature 95 to105 degrees Fahrenheit 35 to 41 degrees Celsius 2 When water reaches the overflow push Tub Fill butt...

Page 9: ...he sprayer into the spa not at the resident Press the trigger on the hand sprayer and adjust the mixing valve if needed to the normal bathing range of 95 to 105 degrees Fahrenheit 35 to 41 degrees Cel...

Page 10: ...rain 5 Press and hold the Disinfect Button located on the control panel As the button is held down the properly mixed cleaning solution is running through the air injection system and out all of the a...

Page 11: ...isibly check that the spa was effectively cleaned during the disinfecting procedure If not repeat the procedure 14 If there is a delay of one or more hours before the next bath we recommend using a to...

Page 12: ...nfectant concentrate from damaging their eyes or skin If disinfectant gets on the skin or in the eyes rinse thoroughly with plenty of water Seek medical advice if irritation occurs 2 Perform the Daily...

Page 13: ...door WARNING If during the safety checks you find parts are missing are excessively worn do not function properly or do not meet the recommended safe operating levels do not operate the equipment unt...

Reviews: