Pendulum CNT-90 Getting Started Manual Download Page 9

Trigger Hold-Off

In this menu you can set the delay during which the stop trigger 

conditions are ignored after the measurement start. A typical use 

is to clean up signals generated by bouncing relay contacts.

Statistics

In this menu you can find the following submenus:

• 

No. of Samples: Set the number of samples used for calculation of 

various statistical measures.

• 

No. of Bins: Set the number of bins used in the histogram mode.

• 

Pacing:  Set  the  delay  between  successive  measurements,  called 

Pacing Time,  to ON or OFF.

• 

Pacing Time: Set the pacing time to a value between 2 µs – 1000 s.

Timebase Reference

Here you can select if the counter is to use the internal or an 

external  timebase  reference.  If  Auto  is  selected,  an  external 

timebase will be used only if it is interpreted as a valid signal, 

i.e.  both  amplitude  and  frequency  must  be  within  specified 

limits. This does not imply, however, that an external reference 

source has to be better in any sense than the internal timebase 

oscillator. The EXT REF indicator at the upper right corner of the 

display shows that the instrument is using an external timebase 

reference.

Miscellaneous

The options in this menu are:

• 

Smart Time Interval: When selected, the counter decides by means of 

timestamping which measurement channel is leading.

• 

Auto Trig LF: In a value input menu you can set the lower frequency 

limit  for  automatic  triggering  and  voltage  measurements  in  the 

range 1 Hz – 100 kHz. A higher limit means faster settling time and 

consequently faster measurements.

• 

Timeout 

 

 

 

 

 

Switch  the  Timeout  function  ON  or  OFF.  In  case  the  input  signal 

gets  interrupted,  the  timeout  system  (if  enabled) will  hold  the  last 

measurement result on screen only during the selected period of time 

(see next paragraph). Then the screen will be blanked, and a pending 

bus query will read a zero result. In case timeout is switched off when 

the signal gets interrupted, the display will freeze, i.e. the result of 

the  last  complete  measurement  will  stay  on  screen  indefinitely.  A 

pending bus query will also wait indefinitely for a response, unless 

the test system controller has enabled its own timeout.

• 

Timeout Time   

 

 

 

 

Set  the  maximum  time  the  instrument  will  wait  for  a  pending 

measurement to finish before outputting a zero result. The range is 

10 ms to 1000 s.

• 

Math/Limit 

 

 

 

 

 

Here you will find the menus for mathematical postprocessing of the 

measurement result and for setting up the limit testing function.

Math

You  can  modify  the  measurement  result  mathematically  by 

scaling  or  offsetting  before  presentation  on  the  display.  This 

feature  can  be  used  for  getting  revolutions/min  instead  of 

Hz or for recalculating the frequency in case a device causing 

frequency conversion (e.g. a multiplier or a mixer) is part of the 

system under test.

Select one of four formulas and enter the constants K, L and M to 

make the counter show directly what you want, without tedious 

recalculations. X stands for the current unmodified measurement 

result. See the User’s Manual for a closer description.

Limits

This  menu  is  used  for  setting  numerical  limits  and  selecting 

the way the instrument will report the measurement results in 

relation to them. See the section Presentation Modes on page 

3-4  for  a  short  description  or  the  User’s  Manual  for  a  more 

detailed description.

User Options

From this menu you can reach a number of functions that are not 

directly involved in the measurement process.

Save/Recall Menu

Twenty complete front panel setups can be stored in nonvolatile 

memory;  the  first  ten  of  them  can  be  user-protected.  The 

different setups can be individually labeled to make it easier for 

the operator to remember the application.
The following can be done:

• 

Save Current Setup: Select one of twenty positions.

• 

Recall Setup: Here you will find a factory-programmed default setup 

as well as any setups you may have stored before.

• 

Modify Labels: The seven soft keys right below the display plus the 

numeric  input  keys  6,  7,  8,  9,  0  are  used  for  entering  lower-case 

letters and digits much in the same way as you write SMS messages 

on a cell phone. Each label can hold seven characters.

9

Summary of Contents for CNT-90

Page 1: ...rch 2020 Timer Counter Analyzer CNT 90 CNT 91 CNT 90 AN Frequency Calibrator Analyzer CNT 91R Microwave Counter Analyzer CNT 90XL Reproduction and distribution of this technical manual is authorized f...

Page 2: ...cription of Keys 3 3 Power 3 3 Select Function 3 3 Autoset Preset 3 3 Move Cursor 3 3 Enter 3 3 Exit OK 3 3 Cancel 3 3 Presentation Modes 3 4 VALUE 3 4 STAT PLOT 3 4 Remote 3 5 Hard Menu Keys 3 5 Inpu...

Page 3: ...al To ensure the correct and safe operation ofthe instrument it is essential that you follow generally accepted safety procedures in addition to the safety precautions specified in this manual The ins...

Page 4: ...available line voltage Fuse The secondary supply voltages are electronically protected against overload or short circuit The primary line voltage side is protected by a fuse located in the power suppl...

Page 5: ...Connector The address is set from the User Options Menu USB Connector A serial interface according to USB 2 0 12 Mb s provides a fast communication link to your PC Optional Main Input Connectors This...

Page 6: ...urown User Interface The fundamental idea has been to facilitate even complex measurements by a consistent interactive user interface that fully exploits the large graphic display The casual as well a...

Page 7: ...ress this key Exit OK This hard key performs the Save Exit operation Youwill confirm your selections by depressing the key and at the same time you will exit the current level and enter the next highe...

Page 8: ...or DC Impedance 50 Q or 1 MQ Attenuation 1x or 10x Trigger Manual or Auto always Auto 10 90 when measuring transition times Trigger Level If Manual has been selected the absolute level can be adjuste...

Page 9: ...measurement will stay on screen indefinitely A pending bus query will also wait indefinitely for a response unless the test system controller has enabled its own timeout Timeout Time Set the maximum t...

Page 10: ...e result will remain on the display until a new measurement is triggered by pressing the RESTART key The HOLD sign in the upper right corner of the screen indicates that no new measurements are taking...

Page 11: ...e measurement itself This is known asoffsetting In the following the direct unscaled result of the counter s measurement process will be referred to as X To set up the counter to display any deviation...

Page 12: ...NEG depending on whether you want to measure the positive or the negative pulse width Finish by selecting the input channel Jitter Measurements Statistics You can make statistical measurements such a...

Page 13: ...Phase Arel B benefit from having fixed trigger levels This is because a change in trigger level causes a change in the measured phase shift and Auto might change the trigger level between measurements...

Page 14: ...ance 1 s Activate the time out function by toggling the soft key Timeout to ON Disconnect the signal and check that the counter shows after one second Keep in mind that the value you select for this t...

Page 15: ...m UNITED STATES phone 1 866 644 1230 us office pendulum instruments com POLAND phone 48 58 681 8901 info pendulum instruments com Pendulum Instruments 2021 March 06 2020 CNT 90 Series rev 1 Specificat...

Reviews: