Pendulum CNT-90 Getting Started Manual Download Page 8

Histogram Mode

Successive measurement results are converted into a histogram.

The  number  of  bins  along  the  horizontal  axis  can  be  set  by  the 

user through the Settings -> Stat menu. The bins in the histogram 

are autoscaled based on the measured data. Limits, if enabled, and 

the running mean value X  are shown as vertical dotted lines. The 

center of the graph is indicated by a filled triangle on the X-axis. The 

corresponding  numerical  value is displayed below it and so is the 

scale factor. Enabled limits affect the autoscaling so as to visualize 

the current measurements and the set limits simultaneously.

Trend Plot Mode

This mode is used for observing periodic fluctuations or possible  

trends. A trend plot terminates (if HOLD is activated) or restarts 

(if  RUN  is  activated)  after  the  set  number  of  samples  has  been 

completed.  The  trend  plot  is  always  autoscaled  based  on  the 

measured  data,  starting  with  0  at  restart.  Limits  are  shown  as 

horizontal lines, if enabled.

Remote

When  the  instrument  is  controlled  from  the  GPIB  bus,  and  the 

remote line is asserted, or when the instrument is controlled from 

the  USB  bus,  then  the  presentation  mode  changes  to  Remote, 

indicated by the label REM on the display. The main measurement 

result and the input settings are displayed in this mode.

Hard Menu Keys

These keys are mainly used for opening fixed menus, from which 

further selections can be made, for instance by means of the soft 

keys.

Input A

By depressing this key, the bottom part of the display will show the 

settings for Input A.
The active settings are in bold characters and can be changed by 

pressing the corresponding soft key below the display. You can also 

move the cursor, indicated by text inversion, with the RIGHT and 

the LEFT arrow keys to the desired position and then make the 

selection by pressing the UP or the DOWN arrow keys. You can 

also use the ENTER key.

The selections that can be made in this menu are:

• 

Trigger  Slope:  positive  or  negative,  indicated  by  corresponding 

symbols

• 

Coupling: AC or DC

• 

Impedance: 50 Q or 1 MQ

• 

Attenuation: 1x or 10x

• 

Trigger:  Manual  or  Auto  (always  Auto  10/90  %  when  measuring 

transition times).

• 

Trigger Level: If Manual has been selected, the absolute level can be 

adjusted with  the  up/down  arrow  keys  or  by  entering  a  numerical 

value from the keyboard.

• 

Filter:  On  or  Off.  Pressing  ENTER  or  the  soft  key  below  the  text 

opens up the Filter Settings menu. See below.

You can select a fixed 100 kHz analog

filter or an adjustable digital filter. The equivalent cutoff frequency 

is set via the value input menu that opens if you select Digital LP 

Frequency from the menu above.

Input B

The available settings under Input B are equal to those available 

under Input A.

Settings

This key accesses a range of more sophisticated instrument settings 

that  usually  need  not  be  changed  for  basic  measurements.  The 

figure below shows the display after recalling the default settings 

via the USER OPT menu.

A detailed description of all the possibilities is far beyond the scope 

of this introductory manual. See the User’s Manual for full details 

on functions mentioned and unmentioned here.

Meas Time

This  value  input  menu  is  only  useful  if  you  select  a  frequency 

function. Longer measuring time means fewer measurements per 

second but gives higher resolution. You can change the Measuring 

Time by entering a numerical value, or you can use the UP/DOWN 

arrow keys to increase or decrease the current value.

Burst

This settings menu facilitates measurements on pulse-modulated 

signals  and  is  only  useful  if  Frequency  Burst  is  the  main 

measurement function.
Both the carrier frequency and the modulating  frequency – also 

known as the pulse repetition frequency (PRF) – can be measured, 

often without the support of an external arming signal (see below).

Arm

Arming is the general term used for the means to control the actual 

start or stop of a measurement. When arming is used, the normal 

free-running  mode  is  inhibited,  and  triggering  takes  place  only 

when certain pretrigger conditions are fulfilled.
The signal(s) used for initiating the arming can be applied to three 

channels,  and  the  start  channel  can  be  different  from  the  stop 

channel. All conditions can be set via the menu below.

8

Summary of Contents for CNT-90

Page 1: ...rch 2020 Timer Counter Analyzer CNT 90 CNT 91 CNT 90 AN Frequency Calibrator Analyzer CNT 91R Microwave Counter Analyzer CNT 90XL Reproduction and distribution of this technical manual is authorized f...

Page 2: ...cription of Keys 3 3 Power 3 3 Select Function 3 3 Autoset Preset 3 3 Move Cursor 3 3 Enter 3 3 Exit OK 3 3 Cancel 3 3 Presentation Modes 3 4 VALUE 3 4 STAT PLOT 3 4 Remote 3 5 Hard Menu Keys 3 5 Inpu...

Page 3: ...al To ensure the correct and safe operation ofthe instrument it is essential that you follow generally accepted safety procedures in addition to the safety precautions specified in this manual The ins...

Page 4: ...available line voltage Fuse The secondary supply voltages are electronically protected against overload or short circuit The primary line voltage side is protected by a fuse located in the power suppl...

Page 5: ...Connector The address is set from the User Options Menu USB Connector A serial interface according to USB 2 0 12 Mb s provides a fast communication link to your PC Optional Main Input Connectors This...

Page 6: ...urown User Interface The fundamental idea has been to facilitate even complex measurements by a consistent interactive user interface that fully exploits the large graphic display The casual as well a...

Page 7: ...ress this key Exit OK This hard key performs the Save Exit operation Youwill confirm your selections by depressing the key and at the same time you will exit the current level and enter the next highe...

Page 8: ...or DC Impedance 50 Q or 1 MQ Attenuation 1x or 10x Trigger Manual or Auto always Auto 10 90 when measuring transition times Trigger Level If Manual has been selected the absolute level can be adjuste...

Page 9: ...measurement will stay on screen indefinitely A pending bus query will also wait indefinitely for a response unless the test system controller has enabled its own timeout Timeout Time Set the maximum t...

Page 10: ...e result will remain on the display until a new measurement is triggered by pressing the RESTART key The HOLD sign in the upper right corner of the screen indicates that no new measurements are taking...

Page 11: ...e measurement itself This is known asoffsetting In the following the direct unscaled result of the counter s measurement process will be referred to as X To set up the counter to display any deviation...

Page 12: ...NEG depending on whether you want to measure the positive or the negative pulse width Finish by selecting the input channel Jitter Measurements Statistics You can make statistical measurements such a...

Page 13: ...Phase Arel B benefit from having fixed trigger levels This is because a change in trigger level causes a change in the measured phase shift and Auto might change the trigger level between measurements...

Page 14: ...ance 1 s Activate the time out function by toggling the soft key Timeout to ON Disconnect the signal and check that the counter shows after one second Keep in mind that the value you select for this t...

Page 15: ...m UNITED STATES phone 1 866 644 1230 us office pendulum instruments com POLAND phone 48 58 681 8901 info pendulum instruments com Pendulum Instruments 2021 March 06 2020 CNT 90 Series rev 1 Specificat...

Reviews: