T
ROUBLESHOOTING
Combination Warmer
▪
Operator’s Manual
▪
M N-47320
▪
Rev 1
▪
1/21
36
*
All non-critical codes can be cleared using the standby button. Critical errors (marked with a *) can
only be cleared by setting the power switch at the rear of the warmer to the OFF position and
allowing the warmer to cool.
*EFAn
Fan or Fan Sensor
Failure
◾
Check to make sure the fan sensor wires did not disconnect from the fan
or the control board.
◾
If the error persists after checking the wires, replace the fan.
Flashing yellow
iv fluid
temperature
status
indicator
Flashing yellow
LED is illuminated
within iv storage
shelf
◾
Set the power switch to the OFF position.
◾
Remove any iv bags from inside the warmer.
◾
Set the power switch to the ON position.
◾
Visually check to see if flashing yellow LED remains illuminated.
◾
Contact service if error persists.
Code
Refers to
Action required