background image

CAUTION

*Do not inflate the cuff on the samb limb which other monitoring ME equipment is 

applied around simultaneously, because this could cause temporary loss of function of 

those simultaneously-used monitoring ME equipment.

*On the rare occasion of a fault causing the cuff to remain fully inflated during 

measurement, open the cuff immediately. Prolonged high pressure (cuff pressure 

 

300mmHg or constant pressure 

 15mmHg for more than 3 minutes) applied to the 

wrist may lead to an ecchymosis.

*Please check that operation of the device does not result in prolonged impairment of 

patient blood circulation.

* When measuring, please avoid compression or restriction of the connection tubing.

* The device cannot be used with HF surgical equipment at the same time.

* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER 

was clinically investigated according to the requirements of ISO 81060-2:2013.

* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please 

contact the manufacturer.

* This device is contraindicated for any female who may be suspected of, or is 

pregnant. Besides providing inaccurate readings, the effects of this device on the fetus 

are unknown.

* Too frequent and consecutive measurements could cause disturbances in blood 

circulation and injuries.

* This unit is not suitable for continuous monitoring during medical emergencies or 

operations.Otherwise, the patient’s wrist and fingers will become anaesthetic, swollen 

and even purple due to a lack of blood.

* When not in use, store the device in a dry room and protect it against extreme 

moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on the 

storage case.

* This device may be used only for the purpose described in this booklet. The 

manufacturer cannot be held liable for damage caused by incorrect application.

*This device comprises sensitive components and must be treated with caution. 

Observe the storage and operating conditions described in this booklet.

* The maximum temperature that the applied part can be achieved is 42.5

 while the 

environmental temperature is 40

.

* The equipment is not AP/APG equipment and not suitable for use in the presence of a 

flammable anesthetic mixture with air of with oxygen or nitrous oxide.

* Warning: No servicing/maintenance while the ME equipment is in use.

* The patient is an intended operator.

* The patient can measure data and change battery under normal circumstances and 

maintain the device and its accessories according to the user manual.

* To avoid measurement errors, please avoid the condition of strong electromagnetic 

field radiated interference signal or electrical fast transient/burst signal.

* The blood pressure monitor and the cuff are suitable for use within the patient 

environment. If you are allergic to polyester, nylon or plastic, please don't use this 

device. 

* During use, the patient will be in contact with the cuff. The materials of the cuff have 

been tested and found to comply with requirements of ISO 10993-5:2009 and ISO 

10993-10:2010. It will not cause any potential sensization or irritation reaction.

* If you experience discomfort during a measurement, such as pain in the wrist or other 

complaints, press the START/STOP button to release the air immediately from the cuff. 

Loosen the cuff and remove it from your wrist.

* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. 

Should the cuff not deflate when pressures reaches 40 kPa (300 mmHg), detach the 

cuff from the wrist and press the START/STOP button to stop inflation.

* Before use, make sure the device functions safely and is in proper working condition. 

Check the device, do not use the device if it is damaged in any way. The continuous 

use of a damaged unit may cause injury, improper results, or serious danger.

* Do not wash the cuff in a washing machine or dishwasher!

* The service life of the cuff may vary by the frequency of washing, skin condition, and 

storage state. The typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after 

maintenance and repair, by retesting at least the requirements in limits of the error of 

the cuff pressure indication and air leakage (testing at least at 50mmHg and 

200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT 

according to the local guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, 

descriptions, calibration instructions,etc., to assist to service personnel in parts repair.

* The operator shall not touch output of batteries and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the 

soft cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile 

cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, 

please contact the SERVICE PERSONNEL of Pearl. Don’t open or repair the device by 

yourself in the event of malfunctions. The device must only be serviced, repaired and 

opened by individuals at authorized sales/service centers.

* Please report to Pearl if any unexpected operation or events occur.

* Keep the unit out of reach of infants, young children or pets to avoid inhalation or 

swallowing of small parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive 

length.

* At least 30 min required for ME equipment to warm from the minimum storage 

temperature between uses until it is ready for intended use. At least 30 min required for 

ME equipment to cool from the maximum storage temperature between uses until it is 

ready for intended use.

* This equipment needs to be installed and put into service in accordance with the 

information provided in the ACCOMPANYING DOCUMENTS;

* Wireless communications equipment such as wireless home network devices, mobile 

phones, cordless telephones and their base stations, and walkie-talkies can affect this 

equipment and should be kept at least a distance d away from the equipment. The 

distance d is caculated by the MANUFACTURER from the 800 MHz to 2.5 GHz column 

of Table 6 of IEC 60601-1-2:2007, as appropriate.

* Please use ACCESSORIES and detachable partes specified/ authorised by 

MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the 

user/patients.

* There is no luer lock connectors are used in the construction of tubing, there is a 

possibility that they might be inadvertently connected to intravascular fluid systems, 

allowing air to be pumped into a blood vessel.

* Please use the device under the environment which was provided in the user manual. 

Otherwise, the performance and lifetime of the device will be impacted and reduced.

CAUTION

4

INTRODUCTION

INTRODUCTION

5

Summary of Contents for Vena TMB-988

Page 1: ...oosing the Vena Blood Pressure Monitor To use the monitor correctly and safely please read the manual thoroughly Please keep this manual well in order to reference in future AVG SYS kPammHg AVG DIA PM...

Page 2: ...BLOOD PRESSURE What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why does my blood pressure fluctuate throughout the day Why do I get a different bl...

Page 3: ...ssure of children Ask your doctor before using it on older children The device is not intended for patient transport outside a healthcare facility The device is not intended for public use This device...

Page 4: ...essure reaches 40 kPa 300 mmHg the unit will automatically deflate Should the cuff not deflate when pressures reaches 40 kPa 300 mmHg detach the cuff from the wrist and press the START STOP button to...

Page 5: ...an inaccurate measurement Blood pressure monitor is detecting an irregular heartbeat during measurement Irregular heartbeat Blood pressure level indicator Indicate the blood pressure level Heartbeat...

Page 6: ...your local recycling guidelines Install the batteries as indicated in the battery compartment Always use the correct battery type 2 AAA batteries Setting Date Time and Measurement Unit 2 Press MEM bu...

Page 7: ...rtably with legs uncrossed feet flat on the floor back and wrist supported Rest for 5 minutes before measuring Wait at least 3 minutes between measurements This allows your blood circulation to recove...

Page 8: ...ent record Press MEM button or SET button to get the record you want The most recent record 1 is shown first Each new measurement is assigned to the first 1 record All other records are pushed back on...

Page 9: ...ut deleting any records press START STOP button before pressing SET button to confirm any delete commands 2 Press SET button to confirm deleting the LCD will display dEL dOnE and then turn off Tips fo...

Page 10: ...en we recommend you seek medical advice Please note that the device does not replace a cardiac examination but serves to detect pulse irregularities at an early stage The chart on the right is the sta...

Page 11: ...rther assistance Refer to the warranty for contact information and return instructions The measurement process does not detect the pulse signal Loosen the clothing on the wrist and then measure again...

Page 12: ...on Type BF applied part Protection against ingress of water Software version Device classification V01 Internally Powered ME Equipment WARNING No modification of this equipment is allowed Temperature...

Page 13: ...evaluation of medical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labeling and inform...

Page 14: ...claration electromagnetic immunity IMMUNITY test Not applicable 3 V m Compliance level IEC 60601 TEST LEVEL Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 Vrms 150 kHz...

Page 15: ...device as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter...

Reviews: