4.3.3. GPIB Input Command Definition
GPIB input commands are ASCII strings used by the GPIB controller for programming GPIB. When
the commands are executed programmatically by the controller,
PeakTech
®
1100 performs special
functions requested by the controller.
4.3.4. GPIB Input Command Types
There are 3 types of GPIB input commands; Parameter, Enumerated, and Direct commands. These
are described in the following sections.
4.3.4.1. Parameter Commands
Parameter commands allow the operational parameter setting of
PeakTech
®
1100 (e.g., Frequency,
Levels, etc.)
Parameter command format:
<header> <numeric argument> <unit> <terminator>
A.
Header of the parameter command is described in Table 4-1.
Table 4-1. Headers for Parameter Command
HEADER
DESCRIPTION
PI
Pilot
MS
Stereo Modulation
FR
Output Frequency
FS
Frequency Step Interval
LE
Output Level
AM
Internal Amplitude Modulation
FM
Internal Frequency Modulation
ST
Storage of an Instrument
Setting
RC
Recall of an Instrument Setting
LS
Level step
B. Numeric argument is made up of maximum of 7 digit numbers including decimal point (.) plus
negative sign(-) if necessary.
Table 4-2. Numeric Keys for Parameter Commands Arguments
Numeric data
DESCRIPTION
0 ~ 9
Numbers
.
Decimal Point
-
Minus Sign
Table 4-3. Units for Parameter Command
UNITS
DESCRIPTION
PC
Stereo
Modulator,
Amplitude
Modulation
KZ
Output Frequency
MZ
Output Frequency
DM
Output Level
DU
Output Level
DB
Level step