Page 15
Technical Specification
Environment
Minimum Operating Ambient Temperature
+5°C (41°F)
Maximum Operating Ambient Temperature
+50°C (122°F)
Maximum Altitude
2000m
Maximum Relative Humidity
80% non-condensing
Minimum Storage Temperature
-20°C (-4°F)
Maximum Storage Temperature
+60°C (140°F)
When taken out of storage, the Generator should be allowed to acclimatize at room temperature for a
minimum of 3 hours before operation.
Generator Outlets
Required Inlet Air Pressure
145 / 116 /101 psig (10 / 8 / 7 barg)*
Maximum Inlet Air Pressure
145 psig (10 barg)
Minimum Inlet Flow
Varies depending on model & configuration
Maximum Outlet Pressure
131 / 101 / 87 psig (9 / 7 / 6 barg)*
Maximum Outlet Flow
Varies depending on model & configuration
N
2
Purity
>99.9%
Electrical Requirements
Voltage
230 VAC ±10%
Frequency
50/60 Hz
Current (i-Flow)
2 A
Current (Dryer / Filtration)
2 A
Pollution degree
2
Installation category
Class II
General
Skid Model
1
2
3
4
5
Dimensions mm (inches) W x D x H
2000 x 1400 x 2343 mm
(78.7 x 55.1 x 92.3 inches)
2240 x 1480 x 2460 mm
(88.2 x 58.3 x 96.6 inches)
Shipping weight kg(lbs)
805 (1775)
885 (1951)
985 (2172) 1215 (2677) 1365 (3009)
Summary of Contents for i-Flow N2