IG-Genius1022
Rev. 2A
Page 4 of 8
10.
11.
Select the appropriate mains cable from the fittings kit
and plug the IEC 60320/C19 socket into the mains
input at the rear of the generator.
Plug the mains cable into an
appropriate 230VAC 50/60Hz
single phase power supply
**NOTE**
Do not touch anything inside the generator while the side panels are
removed and the mains power is connected to the unit.
12.
13.
Switch the generator power
ON
at the switch on the front panel.
The voltmeter is located on the underside of the
compressor area. This will measure and display the
mains voltage that is being supplied to the generator.
Note the displayed voltage.
>= 220V
<= 253V
**NOTE**
If the displayed voltage is 219V or less* we would highly
recommend fitting a transformer. This can be ordered directly from
Peak Scientific, Peak part number is...
06-3200
– Dual Tap Transformer 200V – 230V
<= 219V
>= 194V