PEAK 

Scientific Instruments Ltd 

Zero Air Gas Generator 

Instructions for use Manual 

30

th

 March 2005 

 

ZA015 Issue 3

 

 

Document Change History 

 

Issue No 

Changed 

Initials 

Date 

 

 

 

 

Document Created 

SGM 

01/09/98 

USA Technical support number updated 

FAD 

16/11/04 

New style front added 

FAD 

30/03/05 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for ZA015

Page 1: ...as Generator ZA015 ZA180 Peak Scientific Instruments Ltd Fountain Crescent Inchinnan Business Park Inchinnan Renfrewshire PA4 9RE T Europe 44 0 141 812 8100 T USA 1 866 647 1649 sales T USA 1 866 732...

Page 2: ...Generator Instructions for use Manual 30th March 2005 ZA015 Issue 3 2 Document Change History Issue No Changed Initials Date 1 Document Created SGM 01 09 98 2 USA Technical support number updated FAD...

Page 3: ...Issue 3 3 Contents Page No Document History 2 Warranty Statement 5 Introduction 6 Unpacking Installation 6 Electrical Connection 6 Air Connection 7 Principle of Operation 8 Commissioning 9 Routine Mai...

Page 4: ...persons should operate or service this equipment Warning To avoid risk of electrical shock personal injury or death disconnect power before removing any cover of this equipment Caution To avoid risk o...

Page 5: ...he Goods and or any installation and any relevant packaging as and when reasonably required by the Company 5 Subject to the Company being notified of any defect as is referred to in sub clause 2 hereo...

Page 6: ...ould be taken not to obstruct the ventilation holes on the sides of the unit nor on the ZA180 the fan outlet on the top The generator should be placed on a steady and level base Alternatively the unit...

Page 7: ...isoned if it comes into contact with any halogenated hydrocarbons silicone sprays silicone greases phosphorous compounds lead components high sulphur vapours or other catalyst poisons The air supply s...

Page 8: ...isture and particles down to 0 1 micron The air is then passed to the Zero Air catalytic combustion chamber This works on the principle of catalytic oxidation where hydrocarbons from the incoming comp...

Page 9: ...r a period of time around 40 minutes the Catalytic Chamber should have reached the required temperature When this happens the relay on the Temperature Controller will click and the red LED will go out...

Page 10: ...increased levels of Hydrocarbons being detected on the baseline Reference to the following Fault Finder Chart will identify the source of the problem High CH4 Cat Chamber Hot Cold Control Board LED LE...

Page 11: ...5 3 5 7 0 18 0 Heater Watts 250 250 250 450 Fan Watts N A N A N A 9 0 Electrical Requirements 230V 50Hz 110V 60Hz 1 2A 2 3A 1 2A 2 3A 1 2A 2 3A 2 0A 4 5A Dimensions H x W x D cm Inches 51x25x16 20x10...

Page 12: ...2005 ZA015 Issue 3 12 10 System Diagrams Heater Rod Thermocouple Outlet ZAxxx Pneumatic Diagram Inlet Filter Regulator Cat Chamber Cooling Coil Thermocouple Heater Rod Out N L Temperature Controller 4...

Page 13: ...PEAK Scientific Instruments Ltd Zero Air Gas Generator Instructions for use Manual 30th March 2005 ZA015 Issue 3 13 Maintenance Log Model Serial number Work Done Remarks Date Name...

Page 14: ...PEAK Scientific Instruments Ltd Zero Air Gas Generator Instructions for use Manual 30th March 2005 ZA015 Issue 3 14 Notes...