PEAK
Scientific Instruments Ltd
Zero Air Gas Generator
Instructions for use Manual
30
th
March 2005
ZA015 Issue 3
11
9
Technical Specifications
General Details
Minimum Operating Ambient Temperature
5
o
C (41
o
F)
Maximum Operating Ambient Temperature
45
o
C (113
o
F)
Inlet Conditions
(Free of oil and bulk moisture)
Minimum Air Inlet Pressure
0.69 Barg (10 psig)
Maximum Air Inlet Pressure
8.6 Barg (125 psig)
Maximum Hydrocarbon Content
100 ppm (CH
4
)
Start up time for specified hydrocarbon concentration
40 minutes
Physical Details
Model
ZA015
ZA035
ZA070
ZA180
Max Output (Litres/min)
1.5
3.5
7.0
18.0
Heater Watts
250
250
250
450
Fan Watts
N/A
N/A
N/A
9.0
Electrical Requirements (230V
50Hz)
(110V 60Hz)
1.2A
2.3A
1.2A
2.3A
1.2A
2.3A
2.0A
4.5A
Dimensions (H x W x D) cm
Inches
51x25x16
20x10x6.3
51x25x16
20x10x6.3
51x25x16
20x10x6.3
76x25x16
30x10x6.3
Shipping Weight Kg
Lbs
10Kg
22lb
10Kg
22lb
10Kg
22lb
12Kg
26.4lb
Parts List
Model
ZA015
ZA035
ZA070
ZA180
Inlet Filter Element (110V)
02-1031
02-1031
02-1031
02-1031
Catalytic chamber
06-1075
06-1075
06-1075
06-1075
Heater Element
04-1059
04-1059
04-1059
04-1059
Thermocouple
04-1051
04-1051
04-1051
04-1051
Temperature Controller
04-1170
04-1170
04-1170
04-1170
Cooling Fan (240V)
04-1021
04-1021
04-1021
04-1021
Cooling Fan (110V)
04-1022
04-1022
04-1022
04-1022
Transformer (230V Only)
04-4356
04-4356
04-4356
04-4356