Safe State After Repair
To guarantee the generator is in a safe state after a repair, please ensure the following
conditions are met;
• The electricity cable has been correctly refitted.
The mains cable should be plugged into the IEC power inlet located on the rear
panel of the generator. This unit is classified as SAFETY CLASS 1. THIS UNIT MUST
BE EARTHED.
• The system builds pressure correctly.
With the outlet ports closed, pressure will start to build in the internal storage tanks.
In normal operation, both compressors will cycle alternately. Please refer to the flow
chart section of this manual for more detail.