MODEL NUMBER:

 

PDI-MCA24-AC 

Document Number:

 

PD196-412R2

 

 

Assembly Instructions 

Page 6 of 8 

 

PDi Communication Systems, Inc. 

 40 Greenwood Ln 

 Springboro, Ohio 45066 USA 

 

www.pdiarm.com

 an

www.mymedTV.com

 

 Phone 800.628.9870 

 

 

Mount the Column 

1.

 

Set the base on the workbench on its wheels and set the 
brakes to keep the base from rolling. 

2.

 

Turn over the column so it is resting on the handle and the 
bottom end is resting on the base. 

 
3.

 

Remove the Cover Plate shown in the photo by removing 
the four (4) #10-3

2 button head cap screws with a 1/8” 

Hex Wrench.  Do not completely remove from the assembly.  

This step will allow free movement of the Column Plate to 
gain access to connect the cables. 

 

4.

 

Slide the column plate up to allow access to the cables at 
the base of the column. 

5.

 

Connect the coax cable in the base to the coax cable in the 
column. Tighten to 10 in-lbs. torque. 

6.

 

Connect the Molex connectors for the power cord. 

 

 
 
 

 
 
 
 
 
 
 
 
 
 
 

7.

 

Tip the base up and insert the bottom into the pocket in 
the base.  Be sure that the column rests firmly at the 
bottom of the pocket and is not pinching the cables (The 
Column Plate should still be moveable to ensure there is 
no interference.).  

8.

 

Attach the Column to the Base with the (4) M8 x 30 SHCS, 

PD151-003-12

, and Lock washers 

PDIETLWM8

.  

Tighten to 100 in-lbs. torque.  The cart can now be 
positioned upright on its casters. 

9.

 

Slide the Trim Collar down the column so it rests flush 
against the Base. 

10.

 

Attach the Trim Collar to the Base with the (4) Self-
Threading screws, 

PD151-003-13

 (max tightening 

torque is 6in-lbs. 

11.

 

Re-position the Cover Plate that was removed in step 3 
and re-

install with the same screws with a 1/8” hex 

wrench.  Tighten the screws to 20-25 in-lbs. torque. 

Summary of Contents for PDI-MCA24-AC

Page 1: ...ASSEMBLY INSTRUCTIONS Document Number PD196 412R2 mymedTV com TV Mobile Cart PDI MCA24 AC Mobile Cart with Flat time TV Mount Works with PDi 24 A Series Hospital LED HDTV...

Page 2: ...nd then wiping the unit Specific recommendations can be found at www pdiarm com CAUTION To avoid damage to the surface of the television do not use abrasive or chemical cleaning agents Oxygen Environm...

Page 3: ...sert the plug into the outlet contact your electrician to replace your obsolete outlet 10 Protect the power cord from being walked on or pinched particularly at plugs convenience receptacles and the p...

Page 4: ...eil y compris les amplificateurs qui produit de la chaleur 9 NE PAS neutraliser le dispositif de s curit de la fiche polaris e ou de terre Ce produit est quip d une prise de terre trois fils une fiche...

Page 5: ...ad Zinc Plated 4 each PD197 001 Nut Plate 0 250 20UNC 4 each PD151 003 13 Self Threading Screw 4 each PD151 003 12 Screw M8 x 30mm long Hex Socket Head 4 each PDIETLWM8 Lock Washer M8 External Tooth 4...

Page 6: ...e base of the column 5 Connect the coax cable in the base to the coax cable in the column Tighten to 10 in lbs torque 6 Connect the Molex connectors for the power cord 7 Tip the base up and insert the...

Page 7: ...wn with the four M4 x 10mm screws Mount TV Bracket to Cart NOTE It is recommended that two people work together to mount the TV to the column of the cart One person should hold the TV while the other...

Page 8: ...t to the rear of the column allows the cart to be pushed up closer to a patient in bed Optional Storage Bin 1 Pre assemble the Bin PD180 038 with the screws 2 PDIBHSCS103238Z and Nut Plates 2 PD197 00...

Reviews: