7 8
Peak Hold
The Peak Hold function captures
the peak AC or DC voltage or current.
The meter can capture negative
or positive peaks as fast as
1 millisecond in duration.
1.
Turn the function dial to the A or V position.
2. Press and hold the
PEAK
button until “
CAL
” appears in the
display. This procedure will zero the range selected & meter
go to manual ranging.
3. Press the
PEAK
button,
Pmax
will display.
4. The display will update each time a higher positive peak
occurs.
5. Press the
PEAK
button again,
Pmin
will display. The display
will now update and indicate the lowest negative peak.
6. To return to normal operation, press and hold the
PEAK
button until the
Pmin
or
Pmax
indicator switches off.
Note
: If the Function switch position is changed after a calibration
the Peak Hold calibration must be repeated for the new function
selected.
MAX/MIN
1. Press the
MAX/MIN
key to activate the MAX/MIN recording
mode. The display icon "
MAX or MIN
" will appear. The meter
will go to manual ranging & display and hold the maximum or
minimum reading and will update only when a new “MAX or
MIN” occurs.
2. Press the
MAX/MIN
key and a blinking “
MAX
MIN
” will appear.
The meter will display the present reading, but will continue to
update and store the max and min readings.
To exit MAX/MIN mode press and hold the
MAX/MIN
key for 2 seconds.
Backlight Button
1. Press the BACKLIGHT button to turn the backlighting ON.
2. Press the BACKLIGHT button again to turn the backlighting OFF.