background image

  PDI CA-400D   

 

  AC/DC CURRENT CLAMP 

OPERATOR’S MANUAL 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

International Safety Symbols 

This symbol, adjacent to another symbol or 
terminal, indicates the user must refer to the 
manual for further information. 

SAFETY NOTES

 

 

Do not exceed the maximum allowable input 

range of any function. 

 

Set the function switch OFF when the meter is not 

in use. 

 

INTRODUCTION 

The AC/DC Current Clamp is a transducer which will 

allow your multimeter to measure electrical and/or 

electronic current up to 400  amperes  AC/DC, with a 

frequency response to 50/60Hz. There is no need to 

break a circuit or to remove the insulation when 

measuring current with this clamp. 

When measuring DC current, the CA-400D offers the 

user  “zero” adjustment by turning the 

DC ZERO

 

knob on the clamp probe. 

NOTE

Keep the clamp jaw closed around wire/cable, 

to ensure the accurate readings during current test. 

 

MEASURING CURRENT 

1.

 

Insert the black banana plug into the COM jack 

and the red banana plug into the V  jack of any 

multimeter with a minimum input impedance of 

10k ohms.(OUTPUT and V jack link) 

2.

 

Set the power switch from “OFF” to the desired 

range,  40A (output: 10mV/A) or 400A (1mV/A) 

position. The green LED will light to indicate 

that the clamp is switched on. 

3.

 

For current measurement below 40A, set the unit 

to 40A range and set the multimeter to 400mV or 

600mV AC range for AC current measurements, 

or 400mV  or 600mV DC range for DC current 

measurements. If the measured current exceeds 

40A, set the unit to 400A range. 

4.

 

When  measuring  DC current, always  turn  the 

zero adjustment knob  on the clamp until the 

multimeter reads zero, before taking your 

measurement. 

5.

 

Clamp the jaws around the current-carrying 

conductor and interpret the reading according to 

Step 3 above.  For more detail, refer to section 

below. 

HOW TO INTERPRET READINGS 

When 40A range is selected, the measured current 

value in Amps  if the multimeter reads 100mV, is 

100mV/(10mV/A)=10A.  To simplify that, move the 

decimal point one position to the left. 

When 400A range is selected, the measured current 

value in Amps  if the multimeter reads 100mV, is 

100mV/(1mV/A)=100A.  To simplify that, the mV 

reading is equal to the Amp reading. 

 

APPLICATION NOTES 

1.

 

In the case of DC current, the output is positive 

when the current flows from the upside to the 

underside of the clamp. The red banana plug is 

positive (OUTPUT aperture). 

2.

 

In the case of DC current measurement, a 

hysteresis effect can occur making it  impossible 

to zero the clamp properly. To eliminate this 

effect, open and close the jaws several times then 

turn zero adjustment knob. 

 

Non-Contact AC Voltage Measurements 

 

 

 

WARNING: 

  Risk of Electrocution. Before use, 

always test the Voltage Detector on a known live 

circuit to verify proper operation 

 

Reviews: