![PCS Piezotronics Larson Davis AEC206 Reference Manual Download Page 28](http://html1.mh-extra.com/html/pcs-piezotronics/larson-davis-aec206/larson-davis-aec206_reference-manual_1516419028.webp)
AEC206 Reference Manual
Transfer Functions
A-5
Diffused Field Response
Table A.3 Diffused Field Response
Frequency (Hz)
Response
(dB)
Tolerance
(dB)
100
0.0
±1
125
0.0
±1
160
0.3
±1
200
0.3
±1
250
0.3
±1.3
315
1.0
±1.3
400
1.7
±1.5
500
2.2
±1.5
630
2.9
±1.5
800
3.7
±1.5
1000
4.3
±1.5
1250
5.8
±1.5
1600
8.0
±1.5
2000
11.3
±2
2500
14.4
±3
3150
14.8
±3
4000
13.4
±3
5000
11.1
±3
6300
9.6
±4
800
10.0
±5
10000
7.7
±6