EFP-500 M0001-006 07/26/2011
- 16 -
2.
Connect the Main Channel Assembly and Special Channel Assembly to the quick-
disconnects on the EFP-500. Place these assemblies into the container with the high-
level disinfectant.
3.
Push the Detergent start button. Allow the unit to circulate the high-level disinfectant for
approximately 45 seconds. Disconnect the inlet hose and channel assemblies and
immerse in high-level disinfectant for as long as recommended by HLD manufacturer.
(See bottle label.)
DO NOT RESET THE UNIT.
4.
Fill a separate container with fresh water and place the Inlet Hose and channel
assemblies into the container. Connect the quick-disconnect to the connection marked
“INLET” on the EFP-500. Reattach the Main and Special Channel Assemblies and press
the Water start button to allow the water to circulate for a full cycle, which is
approximately 60 seconds. Disconnect the inlet hose and channel assemblies and
immerse in water. Remove inlet hose and channel assemblies from water. Discard the
rinse water.
5.
Repeat Step 9 for two more times with fresh water (Note: There is no need to disconnect
the inlet hose and channel assemblies for the second and third rinse).
Ignore Water
start button indicator light.
6.
Remove the inlet hose and channel assemblies from the water and place in a clean,
empty container. Press the Air start button to flush air through the EFP-500 and channel
assemblies. Allow to run for a full cycle.
7.
Disconnect the channel assemblies.
8.
Wipe the external surface of the assemblies with a clean, lint-free cloth.
9.
Attach the adapter assemblies to the channel assemblies. Flush with alcohol and then
air.
10.
Store channel assemblies hanging vertically.
8 Flow
Verification
Verifying the flow of the EFP-500 can be easily done and should be checked daily to make sure the
unit is operating to output specifications.
Flow verification is performed without the EFP-500
connected to an endoscope.
Flow Verification of Main Channel Assembly:
1. Fill a container with approximately
three liters
of clean water.
2. Connect the inlet tube assembly to the EFP-500 and place the screened end into the
container with the three liters of water making sure it is fully submerged.
3. Connect the Main Channel Assembly to the EFP-500 and place other ends into the two-liter
graduated container supplied with unit.
Disconnect
the Special Channel Assembly from the
EFP-500.
4. Press the Water Start button and allow the unit to operate for a full cycle making sure the
hose assembly remains in the graduated container.
5. Once the EFP-500 completes its cycle, observe the volume in the graduated container. This
should be
no less than 1.10 liters
.
Flow Verification of Special Channel Assembly:
1. Fill a container with approximately
three liters
of clean water.
Flush with Air
Flush with Water