EFP-500 M0001-006 07/26/2011
- 10 -
Connecting Your Endoscope to the EFP-500
1.
Fill a separate basin/sink with fresh detergent solution per manufacturer’s instructions. The
basin/sink must be large enough to completely submerge the endoscope.
2.
Fill the supplied two-liter container with fresh detergent solution and place it to the side of
the EFP-500
3. Connect the Main Channel Assembly to the appropriate connections on the endoscope. Plug
the other end of the Assembly into the left quick-disconnect on the EFP-500
CAUTION
To prevent detachment of these connections from the
endoscope during operation, make sure the connections are
dry before you insert them. If wet, spray isopropyl alcohol on
both connections and connectors (see below) before insertion
and air dry.
4.
Prior to submerging the endoscope, ensure that the channel plug and water resistant cap
(videoscopes only) are attached to the endoscope. Submerge the endoscope and
connections into the prepared basin/sink.
For those endoscopes with Auxiliary-Water or Elevator-Wire Channels:
Connect the Special Channel Assembly to the appropriate connections on the endoscope.
Then plug the other end of the Assembly into the right side quick-disconnect fitting on the
EFP-500.
5.
Plug the Inlet Hose into the connection marked “INLET” on the EFP-500. Place the end
with the screen into the bottom of the container with detergent.
NOTE:
Always pump fresh, clean solutions through the EFP-500.
Do not place the Inlet Hose in the basin/sink along with the
endoscope. Doing so will circulate contaminated fluids
through the EFP-500.
Air Pipe Connector
Connector Plug
Air Pipe Port
Air & Water Supply
Main Channel Assembly Connections
Main Channel Assembly Connectors
CLEN218