Safety in Practice 

                        EFP-500 

Endo-Flush

 

Endoscope Flushing Pump

Instruction Manual

CLEN218

Summary of Contents for EFP-500

Page 1: ...Safety in Practice EFP 500 Endo Flush Endoscope Flushing Pump Instruction Manual CLEN218 ...

Page 2: ...rica No part of this publication may be reproduced without the express written permission of PCI medical inc PCI medical inc Location 8 Inspiration Lane Chester CT 06412 Mailing PO Box 188 Deep River CT 06417 Tel 860 526 2862 800 862 3394 Fax 866 271 5982 e mail info pcimedical com web www pcimedical com CLEN218 ...

Page 3: ... the EFP 500 10 Channel Selection Process 11 Flushing with Detergent 12 Flushing with Water 13 Flushing with Air 13 Storage 13 5 ROUTINE MAINTENANCE 14 Daily Maintenance 14 Monthly Maintenance 14 6 SYSTEM MAINTENANCE 14 7 DECONTAMINATION CYCLE 14 A EFP 500 Channel Assembly Decontamination 14 B EFP 500 Decontamination 15 8 FLOW VERIFICATION 16 9 QUICK TROUBLESHOOT FAQS 17 Frequently Asked Questions...

Page 4: ...jury It may also be used to alert against unsafe practices or potential equipment damage Protective Grounding NOTE Indicates additional helpful information Environmental Conditions For indoor use only Temperature range 5 C 41 F to 40 C 104 F Main supply voltage fluctuations not to exceed 10 of the nominal voltage Supply Ratings Voltage AC 110 120V 0 7A 50 60 Hz Fuse 2 each Type 5 x 20 mm SloBlo 2 ...

Page 5: ...tructions for manual cleaning When endoscope manufacturer s instructions specify to flush channels with detergent water air using a syringe follow the EFP 500 instructions for automated detergent water air flushing Instruction Manual This instruction manual contains essential information on using this system safely and effectively Before use thoroughly review this manual and the manuals of all equ...

Page 6: ...nd squirting fluid Keep the fluid outputs under water except when checking for flow through the endoscope channels CAUTION Ensure that fluid exiting the distal tip of the endoscope is not pointed in a direction that would allow fluid to contact personnel electrical fixtures or the user CAUTION Do not press the Auxiliary Water or Elevator Wire Select button if not in use This will create back press...

Page 7: ...ing the Suction Instrument Air Water Channels referred to as Main Channel Assembly the Auxiliary Water or Elevator Wire Channels referred to as Special Channel Assembly EFP 500 WB Wall mounting Bracket includes four E Z mount wall anchors and screws EFP 500 PC Power Cord EFP 2L Two liter graduated container EFP 500 AK2 Cleaning Adapter Kit EFP OR O ring replacement pack EFP F5 Screens for Inlet Ho...

Page 8: ...2 Position wall bracket against the wall and mark the position of the four holes 3 Install the four self drilling wall anchors supplied with unit at these points until they are flush with the wall Place bracket over anchors and install using the four Phillips head screws provided 4 For masonry walls use suitable lead or plastic anchors and bolts These anchors are not supplied 5 Place the EFP 500 o...

Page 9: ...tion or the endoscope being placed in an automated endoscope reprocessor AER for disinfection Start Up The unit may be placed on a countertop However it is preferred that the unit be mounted on the wall to prevent inadvertent movement of the device Plug the power cord into the right side of the EFP 500 and into a 115V GFCI protected wall receptacle WARNING Since the EFP 500 is used around water it...

Page 10: ...and air dry 4 Prior to submerging the endoscope ensure that the channel plug and water resistant cap videoscopes only are attached to the endoscope Submerge the endoscope and connections into the prepared basin sink For those endoscopes with Auxiliary Water or Elevator Wire Channels Connect the Special Channel Assembly to the appropriate connections on the endoscope Then plug the other end of the ...

Page 11: ...tor wire channel proceed to the Flushing with Detergent section 2 If the endoscope has an auxiliary or elevator wire channel select the appropriate channel by pressing the corresponding select button on the bottom right of the front panel 3 The LED next to the button will light when the button is selected 4 If the selection is made in error simply press the intended button without pushing the Rese...

Page 12: ... lift the distal tip out of the solution in the basin sink and confirm that fluid is coming out of all channels of the endoscope CAUTION Ensure that fluid exiting the distal tip of the endoscope is not pointed in a direction that would allow fluid to contact personnel electrical fixtures or the user 3 When the cycle time is complete the Detergent LED will stop flashing and stay lit This confirms t...

Page 13: ...nal that the flushing is complete 6 Discard used rinse water Flushing with Air 1 Place the screened end of the inlet hose in an empty container or on the countertop Push the Air Start button The LED indicator light will flash to indicate the air cycle is running The EFP 500 will stop after the time parameters have been met for air flush The LED will stop flashing and remain lit to indicate the air...

Page 14: ...EFP 500 is water resistant and can be cleaned as described above However never immerse the EFP 500 in any fluid CAUTION Never clean the EFP 500 while connected to an electrical outlet After each reprocessing cycle wipe external surface of Channel Assemblies with clean cloth soaked in detergent solution Rinse with clean water 7 Decontamination Cycle A EFP 500 Channel Assembly Decontamination NOTE T...

Page 15: ...l syringe with clean water and flush channel assemblies with clean water Repeat one additional time Detach adapter assemblies and agitate assemblies in water to rinse 7 Remove assemblies from water and place in a clean basin Attach adapter assemblies to channel assemblies and purge with air 8 Fill a sterile syringe with 70 isopropyl or ethyl alcohol 9 Reattach the adapter assemblies and flush chan...

Page 16: ...pty container Press the Air start button to flush air through the EFP 500 and channel assemblies Allow to run for a full cycle 7 Disconnect the channel assemblies 8 Wipe the external surface of the assemblies with a clean lint free cloth 9 Attach the adapter assemblies to the channel assemblies Flush with alcohol and then air 10 Store channel assemblies hanging vertically 8 Flow Verification Verif...

Page 17: ... the unit still does not meet the minimum flow requirements call PCI medical inc 800 862 3394 for assistance 9 Quick Troubleshoot FAQs If your EFP 500 is not running check the following 1 Ensure that the unit is plugged into a live wall outlet Make sure that the GFCI breaker has not been tripped 2 Make sure the On Off switch located on the right side is on 3 In the event that the system is still n...

Page 18: ...tors on the EFP 500 are covered for 90 days If the system is not working correctly or is defective contact PCI medical inc and we will repair or replace the unit at our option during the warranty period PCI medical inc assumes no liability for consequential damages of any kind as a result of the use or misuse of the system by the purchaser the purchaser s employees or any others This warranty is e...

Page 19: ...UC160P OL5 GF UCT160 OL5 GF UE160 AL5 GF UM160 Ultrasound Endoscopes 12 Specifications and Technical Data Electrical 115 V 50 60Hz 0 7A The EFP 500 was tested by Met Laboratories a Nationally Recognized Testing Laboratory NRTL and was certified to UL 61010 1 and CSA C22 2 No 61010 1 safety standards File E112640 Construction Frame Steel Powder Coated Wall bracket Aluminum Dimensions Unit 6 5 Wide ...