Please  note  that  heavy  connecting  cables  should  be  supported  close  to  the  test  object. 
However, an application of force via the cable must be avoided. 
The maximal torque rating of 

1 to 2 Nm

 at the vibration exciter must not be exceeded since this 

may damage the instrument. 
The PCE-VC20 should be placed on a rigid surface. Hand-held operation is not recommended 
because of possible errors. 

6.2 

Calibration

 

After attaching the device under test switch the PCE-

VC20 on by depressing the “On/Off” button 

until the display lights up. Device information, such as hardware and software versions and the 
date  of the last calibration, is displayed for  a few seconds. After short time the  vibration signal 
becomes stable. The display shows the values of vibration frequency and magnitude (Figure 2). 
Please note that these are not measuring value but only normal values. 

 

 

Figure 2: Display in calibration mode 

 
In  the  upper  right  hand  corner  you  will  see  the  actual  accuracy  of  the  vibration  magnitude  in 
percent. After short period of time the percent value should converge on zero. 
If  the  absolute  error  exceeds  3%,  the  percent  value  will  change  to  inverse  characters  and  a 
beep sound will be generated. In this case no calibration can be performed. 
When the maximum weight of the test object is  exceeded, instead of the percent value, an error 
message “OVERL” will appear and the shaker will be switched off. To proceed with calibration, 
first switch the calibrator off. Then reduce the weight of the test object and switch the calibrator 
back on. The admissible weight is up to 500 grams, depending on the selected frequency. The 
overload message can also be eliminated by lowering the vibration magnitude.  
In  the  lower  display  line  the  date  of  the  last  calibration  of  your  PCE-VC20  is  displayed.  This 
entry can only be modified during factory calibration. Please read also section 5 for details about 
calibration. 
The PCE-

VC20 can  be switched  off by depressing the “ON/OFF”  key for  at least  one second. 

Should this be omitted, a timer is provided which switches the calibrator off after 10 minutes. 

Caution: 

The  vibration calibrator is not to  be  used  in  dirty  and  dusty  environments.  Particular 

care  should  be  taken  that  no  ferromagnetic  particles  get  inside  of  the  instrument:  They  may 
damage it within short time. Defects due to dirt and dust are not covered by the warranty. 
 

 

Summary of Contents for PCE-VC20

Page 1: ...ibration Process Calibrator User Manual User manuals in various languages Deutsch fran ais italiano espa ol portugu s nederlands t rk polski can be downloaded here www pce instruments com Last change...

Page 2: ...fety notes 1 2 Specifications 2 3 System description 3 4 Purpose 3 5 Properties 3 6 Operation 4 6 1 Attaching the Test Object 4 6 2 Calibration 5 6 3 Changing the Accumulator 6 6 4 Reset 6 7 Calibrati...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...10 N Max torque 2 Nm Nominal torque 1 Nm Max transverse force 20 Nm Transverse vibration 10 of main axis measured 14 mm above shaker Mounting of test object M5 tapped hole 7 mm deep Clamping magnet A...

Page 5: ...installations Fault finding 5 Properties Handy and robust battery device for laboratory and field use Load independent vibration magnitude with and field use 10 m s 2 vibration acceleration 10 mm s v...

Page 6: ...ransport 6 Operation 6 1 Attaching the Test Object The Vibration exciter of the PCE VC20 has a tapped M5 hole with 7 mm depth for the attachment of the device under test The supplied stud bolts and st...

Page 7: ...If the absolute error exceeds 3 the percent value will change to inverse characters and a beep sound will be generated In this case no calibration can be performed When the maximum weight of the test...

Page 8: ...emory effect Partial charging is permissible Permanent connection of the mains adapter is not recommended This may lead to premature wear of the battery In order to avoid overcharging the battery it i...

Page 9: ...pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devic...

Page 10: ...583 975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english w...

Reviews: