PCE Instruments PCE-VC20 User Manual Download Page 4

 

 

 

©  PCE Instruments 

 

Specifications 

Vibration magnitude (RMS values) 
Vibration acceleration 

10 m/s² 

±3 % at 0 to 40 °C 

Vibration velocity 

10 mm/s 

±3 % at 0 to 40 °C 

Vibration displacement 

10 

μm 

±3 % at 0 to 40 °C 

Vibration frequency 

159.15 Hz 

±0.05 % at -10 to 55 °C 

Radian frequency 

1000/s 

±0.05 % at -10 to 55 °C 

Level indication 

Percent display, above ±3 % beep tone 

Settling time 

< 10 s 

Maximum weight of test 
object for stated accuracy 

600 g 

Vibration exciter 
Dynamic force 

10 N 

Max. torque 

2 Nm 

Nominal torque 

1 Nm 

Max. transverse force 

20 Nm 

Transverse vibration 

< 10 % of main axis, measured 14 mm above shaker 

Mounting of test object 

M5 tapped hole, 7 mm deep 
Clamping magnet 
Adhesive 

Operating temperature 
ranges for 
3% accuracy 
5% accuracy 

 
 
0 to 40 °C 
-10 to 55 °C 

Humidity 

<90 % at 30 °C, no condensation 

Magnetic scatter field at 
shaker 

< 0.2 mT 

Power supply 

Built-in NiMH accumulator, 7.2 V / 1.6 Ah 

Battery operating time 

Approx. 5 h with m = 100 g 

Charge time of accumulator 

3 h 

Charge socket 

DIN 45323 (5.5 / 2.1) 
Positive terminal at center pin 

Charge voltage 

11 to 18 V DC 

Charge current 

<1 A 

Dimensions 

100 x 100 x 120 mm 

Weight 

2.2 kg 

 

Summary of Contents for PCE-VC20

Page 1: ...ibration Process Calibrator User Manual User manuals in various languages Deutsch fran ais italiano espa ol portugu s nederlands t rk polski can be downloaded here www pce instruments com Last change...

Page 2: ...fety notes 1 2 Specifications 2 3 System description 3 4 Purpose 3 5 Properties 3 6 Operation 4 6 1 Attaching the Test Object 4 6 2 Calibration 5 6 3 Changing the Accumulator 6 6 4 Reset 6 7 Calibrati...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...10 N Max torque 2 Nm Nominal torque 1 Nm Max transverse force 20 Nm Transverse vibration 10 of main axis measured 14 mm above shaker Mounting of test object M5 tapped hole 7 mm deep Clamping magnet A...

Page 5: ...installations Fault finding 5 Properties Handy and robust battery device for laboratory and field use Load independent vibration magnitude with and field use 10 m s 2 vibration acceleration 10 mm s v...

Page 6: ...ransport 6 Operation 6 1 Attaching the Test Object The Vibration exciter of the PCE VC20 has a tapped M5 hole with 7 mm depth for the attachment of the device under test The supplied stud bolts and st...

Page 7: ...If the absolute error exceeds 3 the percent value will change to inverse characters and a beep sound will be generated In this case no calibration can be performed When the maximum weight of the test...

Page 8: ...emory effect Partial charging is permissible Permanent connection of the mains adapter is not recommended This may lead to premature wear of the battery In order to avoid overcharging the battery it i...

Page 9: ...pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devic...

Page 10: ...583 975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english w...

Reviews: