Calibration
The instrument is factory calibrated for the specified range, but due to long term drift of
components, re-calibration may be necessary in some cases. For calibrating the instrument, a
reliable source is required. This source should be at least ten times accurate compared to the
range of the instrument.
Note
: PCE Instruments can provide a calibration service that is traceable to international
standards.
We recommend that you return the instrument to the manufacturer or an approved service agent
for calibration. If you use an alternative calibration facility, make sure that it uses these standards.
Battery Info
This page shows the basic battery Information.
•
Battery Level (Percentage)
•
Battery Voltage (in mV)
•
Current (in mA)
•
Battery Status
•
Time to Full (in min)
•
Time to Empty (in min)
About Calibrator
This Page illustrates the Connection diagrams for valid connections to this device.
5
Maintenance and Troubleshooting
5.1
Common Problems
Problem
Possible Causes
Device Not Starting Up
•
Battery Discharged
•
Battery Connection Loose
Reading Fluctuation/
Reading OPEN
•
Wrong / Loose Connections
Error Code on status bar
•
One of the peripheral not working properly.
(Solution: Restart the Device if still error code
showing contact PCE Instruments)
Calibration Out
•
Distortion in due to external noise connection
(Solution: Check connection. If still out, contact PCE
Instruments or Recalibrate Device in authorized
calibration laboratory)
Battery Not Charging
•
Battery Connection Loose
•
Battery Dead