Troubleshooting
Display "OL
“
“OL" appears on the display if the measurement range is exceeded (coating thicker than 1.5
mm). This device can only measure coatings up to 1.5 mm. For thicker coatings, a different
meter will be required.
“OL" also appears if the contact surface of the device was closer than 2 cm to metals. Restart
the meter while keeping a sufficient distance.
Imprecise measured values, no display
Remove the battery and reinsert it after waiting at least 5 seconds. If this does not help or if
inserting a new battery does not help, you can reset the instrument to factory settings. To do this,
switch off the device and then press the key in the middle until AS appears on the display. When
you release the key, the device is reset to factory settings. Then check the measurement accuracy
with the zero standards from the calibration set. If necessary, perform the zero calibration
described in chapter 4.3.
6
Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact
us. You will find the relevant contact information at the end of this user manual.
7
Disposal
For the disposal of batteries in the EU, the 2006/66/EC
directive of the European Parliament
applies. Due to the contained pollutants, batteries must not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use
them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with
your local waste regulations.
If you have any questions, please contact PCE Instruments.