Operation  

4.1 

Insert / replace battery  

- Open the battery compartment cover on the front of the handle (the cover can be unfolded) 
- Connect new 9 V block battery 
- Insert the battery and close the cover.  

If the battery icon indicates a low battery level, the battery must be replaced. A lack of voltage 
affects the stability and reliability of the device. 

4.2 

Make a measurement 

Ensure that the measuring surface is large enough, that the material thickness and, in the case 
of curved objects, the radius of curvature are in accordance with the specifications of the unit. 

Switch on the meter by pressing and holding the key in the middle. When switching on the meter, 
do not place its contact surface on metals - keep at least 2 cm distance from metals. Otherwise, 
"0L" will be displayed while the meter is placed on the metal detected during start-up and incorrect 
readings will be displayed when placed on other surfaces. 

Place the circular contact surface  of the  device  as  perpendicular  as  possible  to the measuring 
location. The surface of the sensor which is mounted on a spring should rest on the test surface 
evenly. Apply the trigger. A short beep sounds and the measured value appears in the display.  

4.3 

Zero calibration  

It  is  recommended  to  perform  a  zero  calibration  regularly  to  increase  the  accuracy  of  the 
measurements.  Prepare  an  uncoated  sample  (one  each  for  ferromagnetic  (Fe)  and  non-
ferromagnetic (N) metals) or use the zero standards supplied. 
With the meter switched on, press the right key until you hear a double beep and the text "ZERO" 
appears and flashes on the display. Place the sensor contact surface of the  tester vertically on 
the sample and apply the trigger. It beeps briefly and the word "ZERO" appears in the display. 
Briefly press the right key to finish the calibration process.  

4.4 

Settings  

Display backlight 

Pressing and releasing the middle key switches the display backlight on or off.  
 

Change units 

Press and release the left key to switch between the units µm, mm, mils. 
 

Display and delete statistical values 

Press  and  release  the  right  key  to  switch  between  the  display  of  the  MEAN  value,  NO  count, 
minimum value MIN and maximum value MAX. The instrument can save 80 values. When you 
continue measuring, the oldest value is overwritten. After switching off the meter, the values are 
deleted. You can also delete the values by pressing and holding the left key. 
 

Auto Power Off 

The unit switches off if no key has been pressed for 3 minutes or if the battery is low. An acoustic 
signal  sounds  before  the  unit  is  switched  off.  If  you  press  the  trigger  for  a  new  measurement 
immediately after the sound or press one of the keys briefly, the statistical data is retained.  
 

 

Summary of Contents for PCE-CT 25FN

Page 1: ...ing Thickness Gauge User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last ch...

Page 2: ...Technical specifications 2 2 2 Delivery scope 2 3 System description 3 3 1 Device 3 3 2 Display and function keys 4 4 Operation 5 4 1 Insert replace battery 5 4 2 Make a measurement 5 4 3 Zero calibra...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...ter 20 mm Material thickness of substrate 0 30 mm 0 05 mm Radius convex 5 mm Radius concave 25 mm Measuring frequency max 2 measurements per second Power supply 9 V block battery Operating conditions...

Page 5: ...PCE Instruments 3 3 System description 3 1 Device 1 Display 2 Keypad 1 Sensor with contact surface 2 Trigger 3 Battery compartment cover...

Page 6: ...omatic power off 5 Current unit of measurement 6 Measured value 7 Statistical value for 2 8 Automatic sensor selection Key functions Key Press and release Press and hold F1 Change m mm mils Deleting t...

Page 7: ...mended to perform a zero calibration regularly to increase the accuracy of the measurements Prepare an uncoated sample one each for ferromagnetic Fe and non ferromagnetic N metals or use the zero stan...

Page 8: ...tings Then check the measurement accuracy with the zero standards from the calibration set If necessary perform the zero calibration described in chapter 4 3 6 Contact If you have any questions sugges...

Page 9: ...tanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co uk...

Reviews: