5. CALIBRATION 

When to calibrate: 

Calibration may be required when it is initially installed, if the scale/balance is moved to a substantial distance. 
This is necessary because the weight of a mass in one location is not necessarily the same in another location. 
Also, with time and use, mechanical deviations may occur. 

How to calibrate: 

 

1. 

Before entering calibration mode, press 

[ON/OFF]

 to turn the power on for 1minutes.

 

2.  Enter calibration mode 

Press 

[ON/OFF]

 to turn the power off. Press 

[UNIT]

 first, then Press

 [ON/OFF]

, while keeping pressing 

[UNIT]

the display will show the A/D value. 

3.  Calibration by the weight 

Wait for the stable indicator to be displayed, Press 

[SET]

 the display will show 

"00SAVE"

, after 2 seconds, the 

display will return to A/D value. 

4. 

Place the weight (eg.200g for FEH300) on platform.  After the stable indicator is displayed, then press 

[SET]

the display will show "00SAVE" and now calibration in finished.

 

5. 

Press 

[ON/OFF]

, turn off the power, press 

[ON/OFF]

 turn on the power again, place the weight (eg.200g for 

FEH300) on the platform to make sure whether weighing is correct, if not, repeat steps 2-5.

 

6.  Return to weighing mode  

Press 

[ON/OFF]

 to turn the power off. Press 

[ON/OFF]

 again to turn on the power and the scale/balance return 

to the weighing mode. 

6. FEATURES 

z

  Auto shut off (optional).           

    

  Auto calibration 

z

  Auto zero tracking                    

 

 Auto backlight (optional) 

z

  Low batter indication                

 

 g/oz/lb/ozt/gn/dwt/ct/n  

z

  Large LCD (6digits 15 mm high)          

 conversion(optional) 

z

  Large square weighting pan 

z

  Net weight/stability indication 

z

 1/30000&1/15000 

division 

available 

 

7. SPECIFICATIONS 

Model number 

Capacity 

Division 

    PCE-BS 300 

300g 

0.01g 

    PCE-BS 3000 

3000g 

0.1g 

    Platter 

φ

100mm 

    Net/gross weight 

600g 

Operating Temperature 

0-40 (32-104 ) 

 

   

Power source 

6×AA Size Batteries or AC/AC 

Adapter 9V/100mA (optional) 

FUNCTION SETTINGS 

Enter function setting mode. 
Press 

[ON/OFF]

 to turn the power off, Press

 [ZERO]

 first, then press 

[ON/OFF]

, while keeping 

[ZERO]

 

pressed, the display will show 

A_ON

 or 

A_OFF

.

 

 

In this direction will find a vision of the measurement technique: 

 

NOTE:

 "This instrument doesn’t have ATEX protection, so it should not be used in potentially explosive 

atmospheres (powder, flammable gases)."

   

 