53
PCE-
830+
PCE-
3007
Range
(0 – 300A for 3007)
(0 – 120A for 3009)
Resolution Accuracy
1 – 20
th
0.1%
±2% of reading ± 4A
21
st
– 50
th
0.1%
±4% of reading ± 4A
51
st
– 99
th
0.1%
±6% of reading ± 4A
(1 to 99
th
order, min. current at the 50 or 60 Hz, 3000A > True RMS > 300A for 3007,
1200A > True RMS )
PCE-
830+
PCE-
3007
Range
(300 – 3000A for 3007)
(120 – 1200A for 3009)
Resolution Accuracy
1 – 20
th
0.1%
±2% of reading ± 40A
21
st
– 50
th
0.1%
±4% of reading ± 40A
51
st
– 99
th
0.1%
±6% of reading ± 40A
Power Factor
(PF)
PCE-
830+
PCE-
6801 or
PCE-
830+
PCE-
6802
Range Resolution
Accuracy
0.00 – 1.00
0.01
± 0.04
PCE-
830+
PCE-
3007
Accuracy
Range Resolution
> 20V and > 30A
< 20V or < 30A
0.000 – 1.000
0.001
± 0.04
±0.1
Phase Angle
(
Φ
)
PCE-
830+
PCE-
6801 or
PCE-
830+
PCE-
6802
Range Resolution
Accuracy
-180° to 180°
0.1°
± 1°
PCE-
830+3
PCE-3
007
(
Φ
, V > 20V, A > 30A)
Range Resolution
Accuracy
-180° to 180°
0.1°
± 2°
0° to 360°
0.1°
± 2°
Peak Value
of ACV
(peak value > 20V)
or ACA
(peak value:
3&(
830+
3&(
6801> 10% of the