Connect the black voltage test alligator to the neutral line “N”.
Connect the red voltage test alligator to the line “A”.
5. Press POWER key to select (P1, Q1, S1, PF1), (P2, Q2, S2, PF2), (P3, Q3, S3, PF3) and (Pt, Qt, St, PFt)
measured values.
6. Frequency (Hz), Phase Angle (
Θ
), Neutral Line Current (I4) and Power Factor
(pF) measurement :
Hz : Press “Hz” key, PF1 will show “Hz”. Press “PF” key to exit.
Θ
: Press “
Θ
” key, PF1 will show “
ψ
”. Press “PF” key to exit.
I4 : Press “I4” key, PF1 will show “I4”. Press “PF” key to exit.
pF : Press “PF” key.
7. Energy measurement:
Press “ENERGY” key, the “Pt”, “Qt”, “St” and “PFt” or “
ψ
t” annunciator
and energy integrate start time are displayed. The energy integrate value and the current time will be continuous
accumulate.
a). KW displays KWh
b). KVAR displays KVARh c). KVA displays KVAh
Press “ STOP ” key to stop energy accumulated and the HOLD annunciator is displayed.
Press
∝