Manual 

 

Introduction 

Thank you for purchasing a viscometer from PCE Instruments. 
The PCE-RVI 5 viscometer was designed to measure the viscosity of paints, varnishes, adhesives, 
pastes and liquid inks according to the ASTM D562 standard. The readings are displayed in KU (KREBS 
units), g (grams) or cP (centipoises). The viscosity measurements can be performed automatically (by 
using pre-defined settings) or manually. Once you finished a measurement, the results can be sent to a 
printer via the integrated RS 232 interface. This makes the viscometer an ideal device for industrial and 
research applications. 

Safety notes 

Please read this manual carefully and completely before you use the device for the first time. The device 
may only be used by qualified personnel and repaired by PCE Instruments personnel. There is no 
warranty of damages or injuries caused by non-observance of the manual. 
 

 

The device may only be used in approved the temperature range. 

 

The case should only be opened by qualified personnel of PCE Instruments. 

 

The instrument should never be placed with the user interface facing an object (e.g. keyboard 
side on a table). 

 

You should not make any technical changes to the device. 

 

The appliance should only be cleaned with a damp cloth / use only pH-neutral cleaner.

 

 

The viscometer should only be operated by specially trained personnel.

 

 

Do not expose the device or its components (except the spindle) to water or other liquids. Do not 
operate the device with wet hands. 

 

The viscometer must not be operated in areas with explosive or inflammable atmospheres. 

 

The device should only be operated in a controlled electromagnetic environment. Transmitters, 
such as cell phones, should not be used close to the viscometer. 

 

Make sure that the device is levelled properly before taking a measurement. Use a water level to 
check and adjust. 

 
This user's handbook is published by PCE Instruments without any guarantee. 
 
We expressly point to our general guarantee terms which can be found in our general terms of business. 
 
If you have any questions please contact PCE Instruments. 
 

 

Summary of Contents for KREBS PCE-RVI 5

Page 1: ...ents UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce inst...

Page 2: ...t up the viscometer 7 5 1 Level the viscometer 7 5 2 Power connection and printer connection 7 5 3 Attach the spindle 7 5 4 Initial turn on and configuration 7 6 Measuring 9 6 1 Automatic measuring mo...

Page 3: ...emperature range The case should only be opened by qualified personnel of PCE Instruments The instrument should never be placed with the user interface facing an object e g keyboard side on a table Yo...

Page 4: ...gory II Operating conditions 10 40 C 80 non condensing Max altitude 2 000 m above mean sea level Pollution degree 2 Protection class IP 20 Dimensions 325 x 190 x 500 mm Weight 8 5 kg 3 2 Contents of d...

Page 5: ...Manual 5 4 System description Adjustable feet Display unit Engine shaft Spindle Supporting base for adaptors Big adaptor Small adaptor...

Page 6: ...nterface description Power switch mains connection and RS 232 interface are located at the back of the viscometer Start button Stop button Display Up button Down button Enter button Power switch Mains...

Page 7: ...engine shaft Then use the locking screw and fix the spindle by rotating clockwise as seen on the picture below After that gently pull the spindle down until it reaches the lowest position possible an...

Page 8: ...ings First you have to select how the date shall be displayed You can choose between EU mode day month year and US mode month day year Use the Up and Down buttons to select and press enter to confirm...

Page 9: ...The sample wait time settings are selected and flashing when you enter this screen Use the Up and Down buttons to change the value and press Enter to confirm Next the sample measure time setting beco...

Page 10: ...ing mode you can adjust the height of the spindle manually This means that you can use different sample containers or sample volumes which cannot be used in automatic mode To enter manual measuring mo...

Page 11: ...our customer support Viscosity reading is unstable or inaccurate Check if device is levelled properly Check if the temperature of the sample is stable Check the characteristics of the sample to be me...

Page 12: ...13 0 2387 587 118 8 2756 63 50 3 179 138 69 5 548 213 83 8 917 288 93 5 1286 363 101 0 1654 438 107 0 2023 513 113 0 2392 588 118 8 2761 64 50 6 184 139 69 7 553 214 84 0 922 289 93 6 1291 364 101 1 1...

Page 13: ...4954 660 122 8 3115 735 127 1 3484 810 131 7 3853 885 135 4 4222 960 138 4 4591 1035 140 3 4959 661 122 8 3120 736 127 2 3489 811 131 8 3858 886 135 4 4227 961 138 4 4595 1036 140 3 4964 662 122 9 31...

Page 14: ...contact PCE Instruments 9 1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire United Kingdom SO31 4RF By phone 02380 987 035 9 2...

Reviews: