Manual
10
If a printer is connected to the viscometer, you can now press Enter to print the results.
Press the Stop button to return to the working mode selection.
Note: If you use the included 600 ml beaker, we recommend filling it up to the 450 ml mark when working
in automatic mode. Use the big adaptor.
Note: You can stop the measuring procedure at any point by pressing the Stop button. The viscometer
stops at its current position and returns to the working mode selection screen.
Note: If the head of the viscometer is not in its starting position (highest position) when entering automatic
measuring mode, it automatically moves back to its starting position.
6.2
Manual measuring mode
When using the manual measuring mode, you can adjust the height of the spindle manually. This means
that you can use different sample containers or sample volumes which cannot be used in automatic
mode.
To enter manual measuring mode, press the Enter button when in working mode selection.
You get to the following screen:
Now you can use the Up and Down buttons to move the spindle to the desired height.
After that, you can start the measurement by pressing the Start button. If you want to stop the
measurement, just press the Stop button.
Once you are done with the measurement, you can print out the results by pressing the Enter button.
To return to the working mode selection, press the Stop button again.
Note: When entering manual measuring mode, the spindle does not move back to its starting position
automatically. You have to do this manually.
Note: Make sure that the spindle is completely covered by the fluid before starting the measurement.