background image

 

 

 

©  PCE Instruments 

 

Applications 

Pharmaceutical industry

 

Monitoring the number of microorganisms under the laminar flow hood; 

Analysis of the cleanliness of a clean room;  
Testing the effectiveness of sterilization. 

Food and beverage industry

 

Detection of microbial contamination in clean rooms according to the HACCP concept. 

Hospitals

 

Detection  of  pathogenic  microorganisms  in  operating  theatres,  as  well  as  in  the 
pharmacy, store room or ICU. 

Electrical industry

 

Measuring particles and micro-bacterial contamination in clean rooms. 

 

Features 

 

1. 

ISO/DIS 14698-1

 

2. 

Reliable  air  sampling  design,  the  sampling  head  has  up  to  300  0.6mm  micropores 
(standard for 100l / min sampling flow sampler) or 300 0.8mm micropores (standard for 
200l / min sampling flow sampler), so that the collected microorganisms can be evenly 
distributed on the surface of the culture medium, reducing the overlap of dust bacteria 
and improving the accuracy rate; 

3. 

With real-time flow monitoring, automatic adjustment of sampling flow rate to ensure 
the accuracy of sampling volume. 

4. 

High-definition  LCD  touch  screen,  rich  display  content,  user-friendly  menu  settings, 
simple and intuitive operation; 

5. 

The starting time and sampling quantity of the automatic air sampler can be set directly 
through the operation panel

 

6. 

The sampler can be controlled by Bluetooth function on mobile phone

 

NOTE: An App is NOT available! 

7. 

The  calibration  operation  is  simple  and  the  flow  can  be  calibrated  by  adjusting  the 
power of the sampling fan directly through the menu options

 

8. 

Fully  automatic  cycle  sampling  function,  can  achieve  up  to  99  continuous  sampling 
times within 100 hours

 

9. 

It is easy to realize the traceability of equipment operation. It can automatically record 
4000 groups of sampling data (including the set sampling location number, sampling 
group  number, sampling  quantity and  sampling time)  and  query the  sampling  record 
data at any time

 

10.  Built  in  high-capacity  rechargeable  lithium  battery,  precise  display  of  battery  power, 

battery can work continuously for more than 15 hours when fully charged

 

11.  Intelligent charging mode,  in order to effectively extend the service life of the battery 

and to realize long-term continuous sampling of the equipment, it can be connected to 
the power charger all the time during operation

 

12.  The  shell  of  the  whole  machine  is  made  of  high-quality  aviation  aluminium,  and  the 

sampling head can be used for steam sterilization or 75% alcohol wiping sterilization

 

13.  The  standard  universal  culture  dish  is  90  ×  13  ~  15mm,  without  expensive  special 

consumables

 

 

 

Summary of Contents for PCE-AS1

Page 1: ...pler PCE AS1 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 16...

Page 2: ...pplications 2 3 Features 2 4 Specifications 3 4 1 Delivery contents 3 5 Device description 3 6 Operating instructions 5 6 1 Key description 5 6 2 Menu 6 7 Maintenance and precautionary measures 10 8 T...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...ime and sampling quantity of the automatic air sampler can be set directly through the operation panel 6 The sampler can be controlled by Bluetooth function on mobile phone NOTE An App is NOT availabl...

Page 5: ...dio alarm Yes Procedure cycle sampling Up to 99 continuous sampling times within 100 hours Sampling parameter setting 99 groups and 30 points Recording function 4000 sets of sampling data records Batt...

Page 6: ...PCE Instruments 4 1 Cover of the collecting head 2 Specimen mount 3 Pump 4 Collecting head stand 5 Control panel with touch screen 6 Power supply 7 On off switch...

Page 7: ...time is displayed when sampling Bluetooth connection status colour means connected gray means not connected When the buzzer is on no diagonal line means on and a diagonal line indicates off Current b...

Page 8: ...just the selected setting options When a setting item is selected the key is valid Enter the setting menu of the instrument and set the corresponding parameters in the corresponding interface Start sa...

Page 9: ...five groups of sampling data You can press the up and down keys to turn the page You can view the first 100 groups of sampling data within the selected time range directly on the instrument at once I...

Page 10: ...password by using the up and down buttons to the right When the password is correct press the CONFIRM key to enter the flow calibration menu as shown in the second figure When entering the flow calib...

Page 11: ...antity is completed the sampling times will be automatically reduced by 1 and the sampling will stop At the same time the countdown of the sampling interval starts When the countdown of sampling inter...

Page 12: ...measures Do not perform sampling with toxic corrosive gas The device is also not intended for use in acidic alkaline environments It is not allowed in areas where there is fine dust exceeding 0 3 mg...

Page 13: ...dicates that some individual keys do not work The internal computer of the sampler has detected a fault Use the key at the back to turn off the meter Wait 5 seconds until you switch it back on All par...

Page 14: ...Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 05...