Troubleshooting 

Problem 

Possible reason 

Solution 

Display does not switch on / 
remains  dark  even  though 
„On / Off“ key is pressed. 

1. Battery low 
2. Fuse faulty 

1. Connect the charger. 
2. Open the lower part of the 
device and insert a new fuse. 

Pump  does  not  work  when 
„Sample“ key is pressed. 

1. Cable in handle is loose or 
broken 
2. Pump output failed 
3. Pump faulty 

1. Connect the cables in the 
grid correctly. 
2. Restore the pump output.  
3. Replace the pump. 

Flow rate is too low. 

1. Battery low 
2.  Holes  in  the  collecting 
head  are  partly  or  fully 
clogged.  
3.  Dish  excessively  covered 
with medium 

1. Connect the charger. 
2. Clean the clogged holes in 
the collecting head. 
3. 

Spread 

the 

medium 

evenly. 

After  starting  the  device, 
display shows something but 
no clear letters or numbers or 
display  indicates  that  some 
individual keys do not work. 

The internal computer of the 
sampler has detected a fault. 

Use  the  key  at  the  back  to 
turn  off  the  meter.  Wait  5 
seconds  until  you  switch  it 
back  on.  All  parameters  will 
be reset. 

 
 
 
 

Warranty 

You  can  read  our  warranty  terms  in  our  General  Business  Terms  which  you  can  find  here: 

https://www.pce-instruments.com/english/terms

 

10  Disposal 

For  the  disposal  of  batteries  in  the  EU,  the  2006/66/EC

 

directive  of  the  European  Parliament 

applies. Due to the contained pollutants, batteries must not be disposed of as household waste. 
They must be given to collection points designed for that purpose. 
 
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use 
them or give them to a recycling company which disposes of the devices in line with law. 
 
For  countries  outside  the EU,  batteries  and  devices  should  be  disposed  of  in accordance  with 
your local waste regulations. 
 
If you have any questions, please contact PCE Instruments. 
 

 

 

 
 
 
 

 

Summary of Contents for PCE-AS1

Page 1: ...pler PCE AS1 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 16...

Page 2: ...pplications 2 3 Features 2 4 Specifications 3 4 1 Delivery contents 3 5 Device description 3 6 Operating instructions 5 6 1 Key description 5 6 2 Menu 6 7 Maintenance and precautionary measures 10 8 T...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...ime and sampling quantity of the automatic air sampler can be set directly through the operation panel 6 The sampler can be controlled by Bluetooth function on mobile phone NOTE An App is NOT availabl...

Page 5: ...dio alarm Yes Procedure cycle sampling Up to 99 continuous sampling times within 100 hours Sampling parameter setting 99 groups and 30 points Recording function 4000 sets of sampling data records Batt...

Page 6: ...PCE Instruments 4 1 Cover of the collecting head 2 Specimen mount 3 Pump 4 Collecting head stand 5 Control panel with touch screen 6 Power supply 7 On off switch...

Page 7: ...time is displayed when sampling Bluetooth connection status colour means connected gray means not connected When the buzzer is on no diagonal line means on and a diagonal line indicates off Current b...

Page 8: ...just the selected setting options When a setting item is selected the key is valid Enter the setting menu of the instrument and set the corresponding parameters in the corresponding interface Start sa...

Page 9: ...five groups of sampling data You can press the up and down keys to turn the page You can view the first 100 groups of sampling data within the selected time range directly on the instrument at once I...

Page 10: ...password by using the up and down buttons to the right When the password is correct press the CONFIRM key to enter the flow calibration menu as shown in the second figure When entering the flow calib...

Page 11: ...antity is completed the sampling times will be automatically reduced by 1 and the sampling will stop At the same time the countdown of the sampling interval starts When the countdown of sampling inter...

Page 12: ...measures Do not perform sampling with toxic corrosive gas The device is also not intended for use in acidic alkaline environments It is not allowed in areas where there is fine dust exceeding 0 3 mg...

Page 13: ...dicates that some individual keys do not work The internal computer of the sampler has detected a fault Use the key at the back to turn off the meter Wait 5 seconds until you switch it back on All par...

Page 14: ...Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 05...

Reviews: