© PCE Instruments
11
8
Troubleshooting
Problem
Possible reason
Solution
Display does not switch on /
remains dark even though
„On / Off“ key is pressed.
1. Battery low
2. Fuse faulty
1. Connect the charger.
2. Open the lower part of the
device and insert a new fuse.
Pump does not work when
„Sample“ key is pressed.
1. Cable in handle is loose or
broken
2. Pump output failed
3. Pump faulty
1. Connect the cables in the
grid correctly.
2. Restore the pump output.
3. Replace the pump.
Flow rate is too low.
1. Battery low
2. Holes in the collecting
head are partly or fully
clogged.
3. Dish excessively covered
with medium
1. Connect the charger.
2. Clean the clogged holes in
the collecting head.
3.
Spread
the
medium
evenly.
After starting the device,
display shows something but
no clear letters or numbers or
display indicates that some
individual keys do not work.
The internal computer of the
sampler has detected a fault.
Use the key at the back to
turn off the meter. Wait 5
seconds until you switch it
back on. All parameters will
be reset.
9
Warranty
You can read our warranty terms in our General Business Terms which you can find here:
https://www.pce-instruments.com/english/terms
10 Disposal
For the disposal of batteries in the EU, the 2006/66/EC
directive of the European Parliament
applies. Due to the contained pollutants, batteries must not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use
them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with
your local waste regulations.
If you have any questions, please contact PCE Instruments.