8. Preparing a balance to work
1. Take all content out of a package.
2. Place the balance on a stable ground not
affected by mechanical vibrations and
airflows.
3. Level the balance with rotating rear legs
6 so that the air bubble in water-level at the
back of the balance is in the middle.
4. (AD50-AD500) Put a pan base 2 on the
mandrel sticking out of a balance cover 3.
Put a pan 1 on a pan base (pan
φ
150mm is
connected to pan base).
5. (AD10-AD12) Put elements 2 on the
mandrel sticking out of a balance, put a
pan 1 .
6. Plug a feeder to the power socket at the
back of the balance.
Leave the pan empty and plug the feeder to the mains 230V. At the end of self-tests, the
balance displays zero indication and is now ready to work
Moisture in the air may condense on the surface of the balance when
transferred to the warmer environment. In this case leave the balance for at
least 4 hours unplugged from the mains for conditioning to avoid wrong
operating or damage of the balance.