Note

: It is always recommended to carry out a 

two-point calibration for better accuracy. 
• For a two-point calibration, place the probe in 
pH 7.01 (or pH 6.86) buffer. After the calibration 
point has been accepted, the “pH 4.01 USE” 
message appears. The message is held for 
12 seconds, unless a valid buffer is recognized. 
If no valid buffer is recognized, then the WRNG 
message is shown. If a valid buffer (pH 4.01, 
pH 10.01 or pH 9.18) is detected, then the meter 
completes the calibration procedure. When the 
buffer is accepted, the LCD shows the accepted 
value with the “OK 2” message on the secondary 
display. The meter then returns to the normal 
measuring mode. 

Note

: When the calibration procedure is completed, 

the CAL tag is turned on. 

 
To quit calibration and to reset to the default 
values 

• After entering the calibration mode and before 
the first point is accepted, it is possible to quit 
the procedure and return to the last calibration 
data by pressing the ON/OFF/MODE button. The 
secondary LCD displays ESC for 1 second and 
the meter returns to normal mode. 
• To reset to the default values and clear a previous 
calibration, press the SET/HOLD button 
after entering the calibration mode and before 
the first point is accepted. 
The secondary LCD displays CLR for 1 second, 
the meter resets to the default calibration 
and the “CAL” tag on the LCD disappears. 
 

METER SETUP 
 

Setup mode allows the selection of the temperature 
unit and of the pH buffer set. 
To enter the Setup mode, press and hold the ON/ 
OFF/MODE button until CAL on the secondary dis-
play is replaced by TEMP and the current tempera-
ture unit (E.g. TEMP °C). Then: 
 

• 

for °C/°F selection, use the SET/HOLD button. 

  After the temperature unit has been selected, 
  press ON/OFF/MODE to enter the buffer set 
  selection mode; press ON/OFF/MODE twice to 
  return to the normal measuring mode. 
 

• 

to change the calibration buffer set, after setting 

   the temperature unit, the meter will show the cur-  
   rent buffer set: “pH 7.01 BUFF” (for 4.01/7.01/   
   10.01) or “pH 6.86 BUFF” (for 4.01/6.86/9.18).   
   Change the set with the SET/HOLD button, then   
   press ON/OFF/MODE to return to normal mode.

 

PH MEASUREMENT & CALIBRATION 
 

• Make sure the meter has been calibrated before 
  use. 
• If the probe is dry, soak it in 

HI 70300 

storage 

  solution for one hour to reactivate it. 
• Place the tip of the probe into the sample to be 
  tested. 
• Stir briefly and wait until the ¹ symbol on the top 
  left of the LCD is turned off. 
• The LCD shows the pH value (automatically com-  
  pensated for temperature) on the primary LCD,  
  while the secondary LCD shows the temperature 
  of the solution. 
• If measurements are taken in different samples 
  successively, rinse the probe tip thoroughly to eli-  
  minate cross-contamination. After cleaning, rinse 
  the probe tip with some of the sample to be   
  measured. 
 

pH calibration 

For better accuracy, frequent calibration of the 
instrument is recommended. In addition, the 
instrument must be recalibrated whenever: 
a) The pH electrode is replaced. 
b) After testing aggressive chemicals. 
c) Where high accuracy is required. 
d) At least once a month. 
• From normal mode, press and hold the ON/ 
OFF/MODE button until OFF on the secondary 
display is replaced by CAL. Release the button. 
• The LCD enters the calibration mode, displaying 
“pH 7.01 USE” (or “pH 6.86 USE” if the NIST 
buffer set was selected). After 1 second the meter 
activates the automatic buffer recognition feature. 
If a valid buffer is detected, then its value is 
shown on the primary display, and REC appears 
on the secondary LCD. If no valid buffer is 
detected, the meter keeps the USE indication 
active for 12 seconds, and then replaces it with 
WRNG, indicating that the sample being measured
is not a valid buffer. 
• For a single-point calibration with buffers pH 
4.01, 9.18 or 10.01, the meter automatically 
accepts 
the calibration when the reading is stable; 
the meter will show on the primary display the 
accepted buffer, with the message “OK 1” on 
the secondary display, and an audible signal is 
produced. After 1 second the meter automatically 
returns to the normal measuring mode. 
If a single-point calibration with buffers pH 7.01 
or 6.86 is desired, then after the calibration point 
has been accepted the ON/OFF/MODE button 
must be pressed in order to return to the normal 
measuring mode. After the button is 
pressed, the meter shows "7.01" (or "6.86") - 
"OK 1", and an audible signal is produced. After 
1 second, the meter automatically returns to 
the normal measuring mode. 

 

Summary of Contents for HI-99131

Page 1: ...s HI 99131 www warensortiment de PCE Deutschland GmbH Im Langel 4 D 59872 Meschede Deutschland Tel 01805 976 990 Fax 029 03 976 99 29 info warensortiment de www warensortiment de 14 Cent pro Minute aus dem dt Festnetz max 42 Cent pro Minute aus dem dt Mobilfunknetz ...

Page 2: ...ttery compartment Note When the meter detects the absence of a temperature probe at its input the Automatic Temperature Compensation is turned off and the meter uses a default value of 25 C 77 F for the temperature measurement and com pensation In this condition the secondary LCD shows 25 0 C 77 0 F blinking When a probe is connected the meter automatically returns to the ATC mode the ATC tag is t...

Page 3: ...rmal mode PH MEASUREMENT CALIBRATION Make sure the meter has been calibrated before use If the probe is dry soak it in HI 70300 storage solution for one hour to reactivate it Place the tip of the probe into the sample to be tested Stir briefly and wait until the symbol on the top left of the LCD is turned off The LCD shows the pH value automatically com pensated for temperature on the primary LCD ...

Page 4: ... solution 500 mL HI 7007L pH 7 01 buffer solution 500 mL HI 7009L pH 9 18 buffer solution 500 mL HI 7010L pH 10 01 buffer solution 500 mL HI 70300L Storage solution 460 mL HI 7061L Cleaning solution 460 mL HI 721312 Hard carrying case HI 710007 Blue shockproof rubber boot HI 76405 Electrode holder SPEZIFICATIONS Range 2 00 to 16 00 pH 5 0 to 105 0 C 23 0 to 221 0 F Resolution 0 01 pH 0 1 C 0 1 F A...

Reviews: