14
Clear Calibration
User can clear calibration to start new applications. When the values are still out of tolerance, the
user should clear it.
a. Press “Left Key” once to enter menu mode (”Root” directory).
b. Press “Up Key” or “Down Key” to activate “Calibration” item, and press “Left Key” once to
enter.
c. Press “Up Key” or “Down Key” to activate ”Clear All” item (Note: User can decide to clear only
part of them,
please see the menu for details), and press “Left Key” once. A dialog box will be
displayed.
d. Pres
s “Left Key” once to confirm the selection and go back, or “Right Key” to cancel and
back.
Note
:
All operations are limited in current work group, and the other groups will not be affected.
8
Troubleshooting
If the instrument gives
no response and can’t be turned on, you can remove the battery, and a
few minutes later, reinstall the battery, and try again. If the problem persists, please contact
PCE Instruments for help.
The following errors can be remedied by system reset:
-Illogical readings
-
Several of the keys don’t work
System reset:
1. Power off the gauge.
2. Press and hold “ZERO Key” and then power on the gauge.
3. Release the “ZERO Key” until a dialog box of system reset shown on the LCD.
4. Press “Left Key” to confirm system reset. The gauge will restart automatically.
Errata:
1. User can press the “Down Key” to refresh LCD display (Clear error display) and reset probe
(To speed up the probe stability) in measuring mode.
2. If an error is shown in menu mode, the user can re-enter the menu.