Multi-point Calibration 

 
a. 

In measure mode, press “CAL key” once to enter calibration mode. 

b. Zero calibration according to the previous chapter and then lay the calibration foil on an 
    uncoated sample. 
c. Apply the probe and raise it after one beep. A reading will be shown. 

Press “Up Key” or 

    

“Down Key” to increase or decrease the readings or hold the key for changing continuously 

    until the required foil thickness value shown. The LCD will show current calibration point 
    

“Ptn    X ”. The “X” will be flashing. 

d. Apply the probe to the test sample several times, and every time raise the probe, if you press 
    

“CAL Key”, after “beep...beep beep...”, it will establish a mean value for current calibration 

    point. 
e. Press “Left Key” to confirm and end current calibration point, and the “X” will be steady. Or 
    

press “Right Key” to cancel and end current calibration point. 

f. For more calibration points, use another calibration foil and repeat the step c, d and e. 
g. Finished. Press “CAL Key” to exit calibration mode. 
 

Note: 

a. Each group has individual zero calibration and multi-point calibration. 
b. When calibration up to 4 points, user must clear the finished points firstly to restart. 
c. Readings measured before will not be affected by new calibration. 
d. In calibration mode, measuring will not be affected. 
e. Recommended to work in single measure mode when calibrating. 
 

Shot-blasted surfaces 

 
The  physical  nature  of shot-blasted  surfaces  results  in coating  thickness  readings that  are too 
high. The mean thickness over the peaks can be determined as follows. 
 
Method one: 
 
a. The gauge should be calibrated according to 

“Multi-point Calibration”. Use a 

    smooth calibration sample with the same curvature radius and the same substrate as the 
    later measuring sample. 
b. Take approx. 10 readings on the uncoated, shot-blasted sample to produce mean value A. 
c. Take approx. 10 further readings on the coated, shot-blasted test sample to produce mean 
    value B. 
d. The coating thickness is T=(B-

A)±S. The “S” is the greater standard deviation of step b and 

    step c. 
 
Method two: 
 
a. The gauge should be calibrated according to 

“Multi-point Calibration” on the shot 

    blasted sample. 
b. Take approx. 10 readings on the test sample to produce the mean value that used as final 
    coating thickness. 
 

 

Summary of Contents for 5000H

Page 1: ...ing Thickness Gauge User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last ch...

Page 2: ...5 2 Basic Measurement Step 5 5 3 User Interface 6 5 4 Measure Mode Only Support Single Mode 7 5 5 Group Mode 7 5 6 Statistics 7 5 7 Probe Mode 8 5 8 Storage 8 5 9 Recall and Delete Readings 8 5 10 Hi...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...designed for non destructive fast and precise coating thickness measurement The principal applications lie in the field of corrosion protection It is ideal for manufacturers and their customers for o...

Page 5: ...de 2 High and low limit alarm 3 Probe mode AUTO MAG EDDY 4 Measuring readings display 5 Statistics display MIN MAX AVG 6 Auto power off indication 7 USB connecting indication 8 Measuring unit 9 Low ba...

Page 6: ...surement no measurement data storage four data groups automatic measurement data storage of up to 2000 measured values Statistical functions Number of measurements average minimum maximum standard dev...

Page 7: ...tery 2 The gauge will work in factory default settings for the first time including single measure mode AUTO probe mode direct group mode DIR etc 3 When power on if in direct group mode DIR the readin...

Page 8: ...Exit b Switch on off the backlight in measure mode Up Key a Move up or roll up b Increasing c Delete the last readings or whole group Down Key a Move down or roll down b Decreasing c Refresh LCD on t...

Page 9: ...thod of switch group mode a Press Left Key once to enter into menu mode Root directory b Press Up Key or Down Key to activate Options item and press Left Key once to enter in c Press Up Key or Down Ke...

Page 10: ...l group mode GENn readings and statistics will be stored and not be lost when power off In direct group mode DIR readings and statistics will be lost when power off Each group has individual high low...

Page 11: ...en you switch the work group the applied alarm settings will also be switched automatically Set high low limit for the current work group as below a Press Left Key once to enter menu mode Root directo...

Page 12: ...ss Up Key or Down Key to select intended item Press Left Key to confirm it 5 13 Data Download User can use the USB cable to connect the gauge with PC and download stored readings If USB connected the...

Page 13: ...asuring accuracy 6 Factors Affecting Measurement Accuracy User needs to know the factors of affecting measurement accuracy before using the gauge The factors are listed below Factors Principle Magneti...

Page 14: ...recalibrate the gauge according to practical applications 7 Calibration User should read chapter 6 Factors Affecting Measurement Accuracy firstly and must recalibrate the gauge according to measured...

Page 15: ...individual zero calibration and multi point calibration b When calibration up to 4 points user must clear the finished points firstly to restart c Readings measured before will not be affected by new...

Page 16: ...roup and the other groups will not be affected 8 Troubleshooting If the instrument gives no response and can t be turned on you can remove the battery and a few minutes later reinstall the battery and...

Page 17: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 18: ...975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english www...

Reviews: