76
Except the instruction is “PR3”, all others are same to the “PR1?”.
ALMp1: Set Alarm Threshold
Instruction
Parameters
Explanation
ALM
p1: Alarm threshold;
Range: 20~200;
Default: 100
ASCII
A
L
M
100
Hex
41H
4CH
4DH
31H, 30H, 30H
Byte
1
1
1
1~3
Return
ACK / NAK
Example: setting alarm threshold as 100dB.
02 01 43
41 4C 4D 31 30 30
03 32 0D 0A
Return: ACK.
02 01 06 03 06 0D 0A
ALM?: Query the Alarm Threshold Setting
Instruction
Parameters
Explanation
ALM
Query parameter: ?
ASCII
A
L
M
?
Hex
41H
4CH
4DH
3FH
Byte
1
1
1
1
Return
Return alarm threshold
Example: query alarm threshold.
02 01 43
41 4C 4D 3F
03 3C 0D 0A
Return: the current alarm threshold is 100dB.
02 01 41
31 30 30
03 70 0D 0A
ETFp1_p2_p3_p4_p5: Set Extended Function
Instruction
P1
P2
P3
P4
P5
Explanation
ETF
p1:
3Profile
Screen;
0=Disable;
1=Enable
p2:
Statistical
Screen;
0=Disable;
1=Enable
p3: Time
History
Screen;
0=Disable;
1=Enable
p4:
Custom
Screen;
0=Disable;
1=Enable
p5: GPS
Screen;
0=Disable;
1=Enable
ASCII
E
T F
1
1
1
1
1
Hex
45H 54H 46H
31H
31H
31H
31H
31H
PCE-428 / 430 / 432