35
(3) Select
Calibration
in the menu and then press
<Enter>
to enter
By Measurement
.
(4)
Adjust
Cal.Level
in the menu, for example adjust to 93.8dB. There is 5s delay after
press
<Start>
to run calibration.
(5) After the end of calibration, sound level meter will update the calibration factor. Press
<Enter>
to apply the results.
(6) Return to
Main screen
and press
<Start/Stop>
to start the measurement. The current
measurement result will be 93.8dB in this example if calibrator is still working.
PCE-428 / 430 / 432