background image

 

Drawing Number: 21110 
Revision: B 

 

OPERATION MANUAL FOR 

QUARTZ PRESSURE SENSOR 

Series 118 & 119

 

 

 

5.0

 

OPERATION

 

 

Most  test  set-ups  will  include  118  &  119  Series 
sensors  coupled  with  charge  amplifiers,  as  shown  in 
the 

Typical 

Connection 

figure 

below. 

For 

applications where the location of the measurement is 
a long distance from where the data must be read, an 
in-line amplifier can be used to drive the signal.  This 
configuration reduces the amount of noise in the data 
and  can  be  seen  in  the  Long  Distance  Connection 
figure to the right. 

 

It  is  important  to  keep  the  charge  amplifier  input 
cable as short as practical.  Input cable capacitance is 
directly  related  to  cable  length  and  is  a  major  factor 
in the amount of electrical noise at the output of any 
charge amplifier. 

 

Depress the ground button of the charge amplifier and 
adjust  electrical  zero  if  necessary.    Range  the 
amplifier  as  required  to  give  the  necessary  full-scale 
voltage. 

 

For  normal,  drift-free  operation,  switch  the  charge 
amplifier  time  constant  selector  to  “medium”  or 
“short” during use. 

 

Typical Connection 

 
6.0 

POLARITY

 

 

Since  most  charge  amplifiers  are  inverting  the  input 
signal,  118  &  119  Series  are  designed  to  produce  a 
negative output voltage for increasing pressure at the 
diaphragm. 

 

Special positive output versions are available for use 
with  our  non-inverting  ICP

®

  source  follower 

amplifiers. 

 

 
 
 
 
ICP

®

 is a registered trademark of PCB Piezotronics 

 

Long Distance Connection 

 

7.0 

MAINTENANCE 

 

Regular  inspection  of  the  diaphragm,  threads,  and 
connector  of  the  118  &  119  Series  sensors  is  good 
practice.   

 

During use in damp environments or after a period of 
storage, the insulation resistance of these sensors may 
degrade (10

12 

ohms is normal).  The only maintenance 

required  on  the  118  &  119  Series  is  cleaning  of  the 
connector to restore insulation resistance.   

 

To  restore  insulation  resistance,  wipe  the  connector 
end  with  a  clean  cloth  or  paper  towel  dipped  in 
Lenium

®

  or  an  equivalent  solvent,  then  rinse  in 

isopropyl alcohol.  The sensor should then be placed 
in  a  250

F  (120

C)  oven  for  several  hours.    A 

vacuum oven, if available, will speed the process.   

 

Contact  the  factory  for  further  assistance  if  the 
insulation resistance cannot be restored. 
 
 

IN-LINE 

AMPLIFIER 

118 & 119 

SERIES 

LOW-NOISE CABLE 

ICP

® 

POWER SUPPLY 

READOUT 

COAXIAL 

CABLE 

118 & 119 

SERIES 

READOUT 

LOW-NOISE CABLE 

CHARGE 

AMPLIFIER 

COAXIAL 

CABLE 

Summary of Contents for PM119B11

Page 1: ...ssure Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ity regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme shock load or other environmental influence or prior to any critical test PCB Piezotronics maintains an ISO 9001 certified metrology laboratory and offers calibration services which are accredited by A...

Page 3: ...urchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers should direct all inquiries to their local distributor or sales office A complete list of distributors and offices can be found at www pcb com Customers within the United States may contact their local s...

Page 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Page 6: ...these sensors a high natural frequency and linearity 3 0 INSTALLATION With a 3 8 24 M10x1 0 for metric mount mounting thread and flush diaphragm design the 118 119 Series transducers mount directly in existing ports machined for PCB Series 108 109 Unlike conventional diaphragm type sensors the 118 119 Series are pressure sensitive over the entire frontal area Extra care should be exercised to avoi...

Page 7: ...eaks at high pressures 3 3 RECESSED MOUNT The recessed installation is shown in the Recessed Mount figure This type of mounting protects the sensor diaphragm from the effects of high flash temperatures and particle impingement due to blast effects thereby prolonging sensor life The recommended range of passage diameters is 0 090 to 0 125 inch 2 29 to 3 18 mm diameter The limitation in this type of...

Page 8: ...temperature effects are usually longer term and will show up as a baseline shift long after the event to be measured has passed For flush mount installations a silicone rubber coating approximately 0 010 inch thick can be effective GE RTV type 106 is recommended and is available from PCB as Model 065A67 Follow manufacturer s instructions to apply It is best to recess the diaphragm 0 010 inch for t...

Page 9: ...LARITY Since most charge amplifiers are inverting the input signal 118 119 Series are designed to produce a negative output voltage for increasing pressure at the diaphragm Special positive output versions are available for use with our non inverting ICP source follower amplifiers ICP is a registered trademark of PCB Piezotronics Long Distance Connection 7 0 MAINTENANCE Regular inspection of the d...

Page 10: ......

Page 11: ......

Reviews: