Periodic recalibration may be required by plants with 
strict certification and traceability requirements. It is 
always  recommended  that  the  user  has  the  sensor 
recalibrated  periodically,  particularly  if  the  sensor 
has  experienced  a  very  high  shock  level  or  extreme 
temperatures  for  extended  periods  of  time.  Some 
plants  develop  in-house  calibration  capabilities  for 
periodically 

verifying 

the 

performance 

of 

accelerometers.  Products  are  available  that  provide 
a  set  1  g-acceleration  level  at  a  fixed  frequency  for 
quick sensor checking. 
 
For  these  reasons,  it  is  recommended  that  a 
recalibration  cycle  be  established  for  each 
accelerometer. This schedule is unique and is based 
on  a  variety  of  factors,  such  as  extent  of  use, 
environmental  conditions,  accuracy  requirements, 
trend 

information 

obtained 

from 

previous 

calibration 

records, 

contractual 

regulations, 

frequency  of  “cross

-

checking”  against  other 

equipment, manufacturer recommendation, and any 
risk associated with incorrect readings. International 
standards,  such  as  ISO  10012-1,  provide  insight  and 
suggested  methods  for  determining  recalibration 
intervals  for  most  measuring  equipment.  With  the 

above  information  in  mind  and  under  “normal” 

circumstances, a conservative suggestion of 12 to 24 
month  recalibration  cycle  for  most  piezoelectric 
accelerometers is advised. 

 
8.1 SENSOR CALIBRATION 

Accelerometer  recalibration  services  are  typically 

performed  by  IMI’s  internal  metrology  laboratory.  

(Other international and private laboratories are also 
available.)    The  IMI  laboratory  is  certified  to  ISO 
9001,  complies  with  ISO  10012-1  (and  former  MIL-
STD-45662A), and uses equipment directly traceable 
to  NIST.    This  assures  an  accurate  calibration  of 
relevant specifications. 
 
The following includes a broad overview of the Back-
to-Back  Calibration  technique  normally  used  when 
calibrating accelerometers.  This technique provides 
a  quick  and  easy  method  for  determining  the 
sensitivity  of  a  test  accelerometer  over  a  wide 
frequency range. 

 
8.2 BACK-TO-BACK CALIBRATION THEORY

 

Back-to-back  calibration  is  perhaps  the  most 
common  method  for  determining  the  sensitivity  of 

piezoelectric  accelerometers.  This  method  relies  on 
a  simple  comparison  to  a  previously  calibrated 
accelerometer,  typically  referred  to  as  a  reference 
standard, as shown in Figure 12. 
 

Figure 12. 

Reference Standard Accelerometer 

 
These  high-accuracy  devices,  which  are  directly 
traceable  to  a  recognized  standards  laboratory,  are 
designed for stability, as well as configured to accept 
a  test  accelerometer.  By  mounting  a  test 
accelerometer  to  the  reference  standard  and  then 
connecting  this  combination  to  a  suitable  vibration 
source,  it  is  possible  to  vibrate  both  devices  and 
compare  the  data  as  shown  in  Figure  13.  (Test  set-
ups  may  be  automated  and vary,  depending  on  the 
type  and  number  for  accelerometers  being 
calibrated.)  
 

Figure 13. 

Typical Back-to-Back Calibration System 

 
Because  the  acceleration  is  the  same  on  both 
sensors, the ratio of their outputs (V

T

/V

R

)  must  also 

be the ratio of their sensitivities. With the sensitivity 
of  the  reference  standard  (S

R

)  known,  the  exact 

sensitivity  of  the  test  sensor  (S

T

)  is  easily  calculated 

by using the following equation:  

𝑆

𝑇

=   𝑆

𝑅

(

𝑉

𝑇

𝑉

𝑅

)

 

 
By varying the frequency of the vibration, the sensor 
may  be  calibrated  over  its  entire  operating 
frequency  range.  The  typical  response  of  an 
unfiltered  accelerometer  is  shown  in  Figure  14. 

Summary of Contents for IMI SENSORS EX637 Series

Page 1: ...el 25mV g 4 to 1 5kHz side exit 2 pin conn Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 800 959 4464 24 hour SensorLin...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...lated 2 20 mA at 18 to 28 VDC source This power is regulated by a current limiting circuit which provides the constant current excitation required for proper operation of ICP sensors In general batter...

Page 7: ...sential as the reliability and accuracy of any measurement system is no better than that of its weakest link Due to the nature of vibration measurements all sensor cables will ultimately fatigue and f...

Page 8: ...CB Piezotronics This Division is wholly contained in the PCB Piezotronics manufacturing facility at 3425 Walden Avenue Depew New York Same address same manufacturing facility Some of the documentation...

Page 9: ...ON A ECN NUMBER 49766 Operating Guide with Enclosed Warranty Information 3424 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 959 4IMI Piezoelectric IC...

Page 10: ...ply low frequency and high frequency measurements Shear design sensors are preferred because of their inherent insensitivity to adverse environmental influences such as case or base strain and thermal...

Page 11: ...wer gain which is extremely close to unity and independent of input voltage Also the diode can be changed to supply higher currents for driving long cable lengths Constant current diodes as shown in F...

Page 12: ...nics However when operating AC coupled signal conditioners with readout devices having an input impedance of less than one megaohm thelow frequencyrangemaybeaffected 4 1 STANDARD STUD MOUNT This mount...

Page 13: ...derside of the mounting base Firmly press down on the assembly to displace any extra adhesive remaining under the base See Figure7 METHOD 2 Direct Adhesive Mount For restrictions of space or for conve...

Page 14: ...interference or radio frequency interference EMI or RFI To avoid ground loops thereshouldonlybeonegroundin the system It is recommended for permanent installations that the sensor becase isolatedand...

Page 15: ...ate Also any current not used by the cable goes directly to power the internal electronicsandwillcreateheat Thismaycausethesensorto exceed its maximum temperature specification For this reason donotsu...

Page 16: ...er the meter or LED and allow the sensor to power up If a faulty condition is monitored first check all system connections then check the functionality of the cable and signal conditioner If the syste...

Page 17: ...and former MIL STD 45662A and uses equipment directly traceable to NIST This assures an accurate calibration of relevant specifications The following includes a broad overview of the Back to Back Cal...

Page 18: ...A of constant current to extend battery life while continuous monitoring systems offer more current in order to drive longer cables Figure 16 shows a typical system for a sensor with integral electron...

Page 19: ...eive assistance via e mail at imi pcb com or visit our website at www pcb com 11 0 CUSTOMER SERVICE WARRANTY IMI a division of PCB Piezotronics guarantees Total Customer Satisfaction If at any time fo...

Page 20: ......

Page 21: ...O DRAFTING 49059 NO CHANGES WITHOUT APPROVAL SCHEDULE DRAWING NO MODIFICATIONS PERMITTED WITHOUT REFERENCE TO THE NOTIFIED BODY SHIELDS TO BE EARTHED AT BARRIER ENDS THE INSTALLER SHALL INSURE THAT TH...

Page 22: ...n Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN SEE SHEET 1 SCHEDULE DRAWING NO MODIFICATIONS PE...

Page 23: ...UNF 5 MountingTorque 2 to5 ft lb 2 7 to6 8 Nm SensingElement Quartz Quartz SensingGeometry Shear Shear HousingMaterial Stainless Steel Stainless Steel Sealing Hermetic Hermetic 6 Electrical Connector...

Page 24: ...CASE ISOLATED ACCELEROMETER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 12 13 18 KRM 12 13 18 RMB 12 13 18 68326 PCB Piezotronics Inc claims proprietary rights in the information disclosed here...

Page 25: ...tes that the product is subject to the special conditions of use specified in the Schedule to this certificate 9 This EU Type Examination Certificate relates only to the design and construction of the...

Page 26: ...The relevant Essential Health and Safety Requirements EHSR5 have been identified and assessed in lntertek Report 103312845CRT 003a Dated 8 August 2019 This Certificate is forth exclusive use of Intert...

Page 27: ...9 MODEL EX M 637XYYYZ SERIES ETCHING DRAWING Page 1 o 2 This Certificate Is for the exclusive use of Intertek s client and is provided pursuant to the agreement between Intertek and its Client Interte...

Page 28: ...f Intertek s client and is provided pursuant to the agreement between Intertek and its Client lntertek n responsibility and liability are limited to the terms and conditions of the agreement Intertek...

Page 29: ...number it indicates that the product is subject to the special conditions of use specified in the Schedule to this certificate 9 This Type Examination Certificate relates only to the design of the spe...

Page 30: ...C 60079 15 Unused connector must be fitted with an appropriately rated blanking cover b Conditions of Manufacture Routine Tests Dielectric Strength Test Per IEC 60079 15 14 Essential Health and Safety...

Page 31: ...3 ETCHING DRAWING 70897 NR 8 8 19 MODEL EX M 637XYYYZ SERIES ETCHING DRAWING Page lof 2 This Certificate is for the exclusive use of lutertek s client and is provided pursuant to the agreement between...

Page 32: ...of lntertek s client and is provided pursuant to the agreement between Intertek and its Client ietertek s responsibility and liability are limited to the terms and conditions of the agreement Intertek...

Page 33: ...Number This Authorization to Mark is for the exclusive use of Intertek s Client and is provided pursuant to the Certification agreement between Intertek and its Client Intertek s responsibility and li...

Page 34: ...nd Non Incendive Equipment For Use In Hazardous Locations R2016 CSA C22 2 157 1992 Ed 3 GI U2 Nonincendive Electrical Equipment For Use In Class I And II Division 2 And Class III Divisions I And 2 Haz...

Page 35: ...YYYZ EX M 638XYYYZ Optional accessory Type of Protection IntrinsicSafety Ia Non Sparking nA Marking Ex ia IIC T4 Ga Ex nA IIC T4 Gc 1960 5 Ta 5 121 C lECEx ETL 19 0034X Approved for issue on behalf of...

Page 36: ...erational Documents as amended STANDARDS The apparatus and any acceptable variations to it specified in the schedule of this certificate and the identified documents was found to comply with the follo...

Page 37: ...barrier item When used in Zone 2 The Accelerometers provide a voltage output when subjected to mechanical motion They are intended to be fixed installation in the explosive atmosphere and to be suppl...

Page 38: ...98 NR 7 24 2019 SECURITE INTRINSEQUE NON SPARK ARC lECEx Ex Ia lECEx Ex nA EX M 637XYYYZ Series Industrial ICP Accelerometer EX M 638XYYYZ Series Industrial ICP Accelerometer Page 2 of 2 DESCRIPTIVE N...

Page 39: ...ETCHING DRAWING Page 1 of 2 ETCHING DRAWING 70900 NR 8 8 19 MODEL EX M 638XYYYZ SERIES ETCHING DRAWING Page 2 of 2 APPROVAL 70910 NR 7 27 19 INTERCONNECTION DIV 1 ZONE 0 Page 1 of APPROVAL 70910 NR 7...

Page 40: ...INSTRUCTIONS FOR USE 70911 NR 08 01 19 EX M 637XYYYZ and EX M 638XYYYZ Series Page 3 of 3 APPROVAL XTAL 71165 NR 7 24 19 DRAWING Page 1 of 1 Note An is included before the title of documents that are...

Page 41: ...data sheets and where those data sheets refer to this declaration of conformity The data sheets for all ICP model numbers referenced above which include the CE mark on such data sheets and refer to th...

Page 42: ...045c GM29047c GM29029s GM29031s GM29033s GM29044s GM29046s GM29048s I the undersigned hereby declare that the equipment specified above conforms to the above Directive s Standard s Place Depew NY Date...

Page 43: ...erenced above which include the CE mark on such data sheets and refer to this Declaration of Conformity are hereby incorporated by reference into this Declaration Conform to the following UK Legislati...

Page 44: ...045c GM29047c GM29029s GM29031s GM29033s GM29044s GM29046s GM29048s I the undersigned hereby declare that the equipment specified above conforms to the above Directive s Standard s Place Depew NY Date...

Reviews: