background image

Drawing Number: 21102 

Revision: B 

 

 

 

 

OPERATION MANUAL FOR 

3  

Series 138 

 

 

The  steel wire can have a weight attached to it to 

submerge the entire assembly.  Tape or attach the 

coaxial cable loosely to the steel wire at ~one foot 

intervals. Attach a monofilament line (~50 lb. test) to 

the end of a steel wire braid, and let it hang down past 

the end of the sensor. 

 

Attach a three to five lb. weight to the end of the 

monofilament  and  tape  the  sensor  to  the 

monofilament. See Image 1 for reference.  

 

 

                Image 1 – Installation Methods 

 

 

3.0 

OPERATION 

 

It is necessary only to supply the sensor with a 2 to 20 

mA constant current at +20 to +30 VDC through a 

current-regulating diode or equivalent circuit.  

 

Most  PCB  signal  conditioners  have  an  adjustable 

current feature allowing a choice of input currents 

from 2 to 20 mA. 

 

In  general,  for  best  resolution,  choose  the  lower 

current ranges, and for driving long cables (to several 

thousand feet) use the higher current, up to 20 mA 

maximum.  Line-impedance  matching  may  be 

necessary  to  produce  flat-frequency  response  over 

long (hundreds or thousands of feet) cables. 

 

Switch  power  on  and  observe  reading  of  bias 

monitoring  voltmeter  on  front  panel  of  the  signal 

conditioner. 

 

If indicator is in green section of indicator panel, the 

IC amplifier is producing proper bias (+8 to  14 VDC), 

cable connections are normal and the system is ready 

to operate. 

 

If the needle moves to the red area of the  fault monitor 

meter, output is zero and a short is indicated.  Short 

could be located in amplifier, cable connectors,  or 

power unit. 

 

If the pointer moves into the yellow area of the fault 

monitor meter, an open circuit is indicated with full 

power supply voltage.  An open circuit could be the 

result of a faulty amplifier, an open cable or open 

connectors. 

 

Allow the sensor to stabilize for about one minute. A 

signal drift may occur when cable is connected to the 

readout instrument. 

 

This  drift  occurs  during  charging  of  the  coupling 

capacitor in the power unit.  The signal will stabilize 

in several minutes. 

 

4.0 

CALIBRATION 

 

Series 138 are calibrated dynamically using a drop 

weight tester.  For best accuracy, use the calibration 

certificate supplied. Factory recalibration is available 

for a nominal charge. 

 

5.0 

POLARITY 

 

This sensor series produces a positive-going output 

voltage  for  increasing  pressure  input.  See  the 

specification sheet for polarity of the sensor. 

 

6.0 

LOW-FREQUENCY RESPONSE 

 

The  low-frequency response  of  an  ICP

®

  system is 

determined by: 

 

l.)  The discharge time constant of the sensors. 

 

2.)  If  AC-coupled  at  the  power   unit,   the coupling 

time constant. 

 

Summary of Contents for ICP W138A02

Page 1: ...Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail i...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...ducted over a single conductor coaxial cable with a shield serving as signal return Special low noise cables are not required Series 138 is used by the military for underwater explosive testing commer...

Page 7: ...high shock can flex the cable connector joint enough to move the center pin of the 10 32 cable connector This can cause noise on the output signal Protecting the connection from shock may be required...

Page 8: ...itioner If indicator is in green section of indicator panel the IC amplifier is producing proper bias 8 to 14 VDC cable connections are normal and the system is ready to operate If the needle moves to...

Page 9: ...sor l Keep connections clean and dry 2 Do not exceed specified maximum range 3 Do not subject sensor to temperaturesexceeding 100 F 4 Do not apply voltage to sensor withoutcurrent limiting diodes or o...

Page 10: ...nt Tourmaline Tourmaline HousingMaterial Stainless Steel Stainless Steel Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 61 oz 17 0 gm 4 All specifications are at room temperature...

Page 11: ......

Reviews: