INSTRUCTIONS FOR USE - EX378XYYY  Microphone System Series 

 

Model(s) 

EX378XYYY Series where “X” is any letter A-Z, “YYY” is up to three numbers 0-999 

Markings 

PCB 
Depew, NY 
LCIE 17 ATEX 3059 X 
Ex I M1, Ex ia I Ma 
Ex II 1G, Ex ia IIC T4 Ga 
IECEx LCIE 17.0063X 
Ex ia I Ma 
Ex ia IIC T4 Ga 
ETL c/us Listed Intertek 5010230 
CL I, DIV 1, GR A,B,C,D 
CL I, ZONE 0, AEx/Ex ia IIC T4 Ga 
-40

C ≤ T

amb

 ≤+80

Putting Into Service 

Powering:  All ICP

® 

Microphone Systems require constant current excitation for proper operation.  For 

this  reason,  use  only  PCB  constant-current  signal  conditioners  or  other  approved  constant-current 
sources.    The  power  supply  consists  of  a  current-regulated,  18  to  30  VDC  source.    This  power  is 
regulated  by  a  current-limiting  circuit,  which  provides  the  constant-current  excitation  required  for 
proper operation of ICP

 Microphone Systems. 

In general, battery-powered devices offer versatility for portable, low-noise measurements, whereas 
line-powered units provide the capability for continuous monitoring.  Consult the Vibration Division’s 
product catalog for more information about signal conditioners. 
 

NOTE: Under no circumstances should a voltage be supplied to an ICP

 Microphone System without 

a current-regulating diode or equivalent electrical circuit.  This may include ohmmeters, multi-meters 
and continuity testers. 

Safe Use 

After completing the system setup, switch on the signal conditioner and allow 1 to 2 minutes for the 
system to stabilize.  The meter (or LED) on the signal conditioner should be reading “green.”  This 
indicates proper operation and you may begin taking measurements.  If a faulty condition is indicated 
(red or yellow reading), first check all system connections, then check the functionality of the cable 
and  signal  conditioner.    If  the  system  still  does  not  operate  properly,  consult  a  PCB  factory 
representative. 
 

NOTE:  Always  operate  the  Microphone  System  within  the  limitations  listed  on  the  enclosed 
Specification  Sheet.    Operating  the  device  outside  these  parameters  can  cause  temporary  or 
permanent damage to the Microphone System. 

Assembling 

The EX378XYYY Series come assembled in system consisting of a Microphone and a Pre-
Amplifier. It does not require any assembly.  Only mounting to low profile stands and clamps while 
taking measurements.  

Dismantling 

Other than removal from the mounting, there is no disassembly of the sensor required to take it out of 
service. 

Maintenance 

Routine maintenance, such as the cleaning of electrical connectors, housings, and mounting surfaces 
with solutions and techniques that will not harm the physical material of construction, is acceptable.  

Summary of Contents for EX378B02

Page 1: ...ystem Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mai...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...in which the instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument St...

Page 7: ...owered with a 24V or a 12V phantom power supply but this will limit the maximum output voltage Externally Polarized Microphone System Prepolarized Microphone System Phantom Powered Microphone System T...

Page 8: ...em such as the preamplifier signal conditioner or instrumentation Conversely the Cartridge Thermal Noise CTN specification provides the lowest measurable sound pressure level that can be detected abov...

Page 9: ...rugged grid cap shown below and as an independent part of a rugged kit to extend the lifespan of the precision measuring device caution should be taken to ensure the delicate diaphragm sensing element...

Page 10: ...ustry standards ensures accurate calibra tion against relevant specifications We also carry a line of acoustic calibrators that can be used to calibrate microphone sensitivity on site as needed Hand c...

Page 11: ...ow 1 to 2 minutes for the system to stabilize The meter or LED on the signal conditioner should be reading green This indicates proper operation and you may begin taking measurements If a faulty condi...

Page 12: ...outine calibration of sensors by the manufacturer is recommended as this helps build confidence in measurement accuracy and acquired data Danger Areas for pressure relief devices N A not a pressure re...

Page 13: ...14043 3425 Walden Ave Depew New York 14043 No 65238 No 3 of 3 Rev NR ECO 45769 Note Literature such as the manual or marketing materials describing the equipment or protective system must not contrad...

Page 14: ...d a voltage be supplied to an ICP Microphone System without a current regulating diode or equivalent electrical circuit This may include ohmmeters multi meters and continuity testers Safe Use After co...

Page 15: ...highly recommended to keep spare cables on hand to enable continuation of the test in the event of a cable failure Install Microphone system per drawing 68216 Adjustment The EX378XYYY series come ass...

Page 16: ...IV 1 GROUPS A B C D CODE IDENT NO 52681 DWG NO SCALE SHEET DRAWN CHECKED ENGINEER TITLE UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES DECIMALS X 05 XX 01 XXX 005 ANGLES 2 DEGREES FILLETS AND RAD...

Page 17: ...oltage 7 Vpk 7 Vpk Output Impedance 50 Ohm 50 Ohm Physical Housing Material Stainless Alloy Stainless Alloy Venting Rear Rear 4 Electrical Connector BNC Jack BNC Jack Mounting Thread Grid 0 5 60 UNS 0...

Page 18: ...MICROPHONE DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 3 1 18 KRM 3 1 18 GGS 3 1 18 66401 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any r...

Page 19: ...ia IIC T4 Ga I M1 Ex ia I CODE IDENT NO 52681 DWG NO SCALE SHEET DRAWN CHECKED ENGINEER TITLE UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES DECIMALS X 05 XX 01 XXX 005 ANGLES 2 DEGREES FILLETS...

Page 20: ...stem Type Ex378XYYY Optional accessory Type of Protection Ex ia Marking Ex ia I Ma Ex ia IIC T4 Ga Approved for issue on behalf of the lECEx Certification Body Position Signature for printed version D...

Page 21: ...ns as set out in lECEx Scheme Rules lECEx 02 and Operational Documents as amended STANDARDS The electrical apparatus and any acceptable variations to it specified in the schedule of this certificate a...

Page 22: ...e the interface to an external cable assembly The equipment is intended to be used for leak detection and acoustic noise tests in gaseous or mining explosive areas SPECIFIC CONDITIONS OF USE YES as sh...

Page 23: ...trinsic safety 1 EN 61326 1 2013 Electrical Equipment for Measurement Control and Laboratory Use EMC EN 61326 2 3 2013 Electrical Equipment for Measurement Control and Laboratory Use EMC EN 61010 1 20...

Page 24: ...venue du G n ral Leclerc FR 92260 Fontenay aux Roses Tel 33 1 40 95 60 60 Fax 33 1 40 95 86 56 I the undersigned hereby declare that the equipment specified above conforms to the a W Direce Standard s...

Page 25: ...nfidentiel s No report s N 144457 692709 8 Le respect des Exigences Essentielles de S curit et de Compliance with the Essential Health and Safety Sante est assure par la conformit a Requirements has b...

Page 26: ...nd a preamplifier The delivered signal from lOP Microphone Systems provides a typical 1 7V peak voltage output signal when making acoustic measurements The microphone cartridge is the sensing element...

Page 27: ...r metallic body and cable shield are reliably connected to the system earth The apparatus must be connected according to instruction manual For final installation the user shall take all necessary pre...

Reviews: