Repair 

In the event that equipment becomes damaged or ceases to operate, arrangements should be made to 
return the equipment to PCB Piezotronics for repair. User servicing or repair is not recommended 
and, if attempted, may void the factory warranty. 
 

Installation 

Overview:  The Microphone System must be mounted on low profile stands and clamps in order to be 
put into service.  Characteristics like location, ruggedness, amplitude range, accessibility, temperature, 
and  portability  are  extremely  critical.  Cabling:    Care  and  attention  to  cable  installation  and  cable 
condition is essential as the reliability and accuracy of any measurement system is no better than that 
of  its  weakest  link.    Due  to  the  nature  of  vibration  measurements,  all  sensor  cables  will  ultimately 
fatigue  and  fail.    Good  installation  practice  will  extend  the  life  of  a  cable,  however,  it  is  highly 
recommended to keep spare cables on hand to enable continuation of the test in the event of a cable 
failure. 
Install Microphone system per drawing 65241. 

Adjustment 

The EX378XYYY series come assembled in a system and no user adjustments are possible.  
However, routine calibration of sensors by the manufacturer is recommended as this helps build 
confidence in measurement accuracy and acquired data. 

Danger Areas (for 
pressure-relief devices) 

N/A – not a pressure relief device. 

Training Instructions 

Industrial sensors must be installed in Hazardous Locations by trained professionals according to 
EN/IEC 60079-14 requirements. 

Details on Safety of 
Protection Category 

Ex ia is “intrinsic safety”, which limits the energy of sparks and surface temperatures to safe levels 
 

Entity Parameters and 
Limits (Values) 

Temperature Range:  -40

C to +80

Ui :30V, Ii :100mA, Pi : 750mW, Ci: 0, Li: 0.36µH 
 

Special Conditions of 
Use 

Version Ex ia : 
The apparatus must only be connected to a certified associated intrinsically safe equipment. This 
combination must be compatible regarding intrinsic safety rules (see electrical parameters). The 
apparatus shall be connected according to drawing 65241. 
 
The mounting of the apparatus into an installation must be carried out in such a way that sensor 
metallic body and cable shield are reliably connected to the system earth. 
 
For final installation, the user shall take all necessary precautions to  maintain the minimum degree 
of protection IP20 of the sensor connection when connected according to the requirements of  
EN 60079-14 and IEC 60079-14 standards.  

Essential 
Characteristics of tools 
fitted to the system (if 
any). 

N/A – No tools are fitted to the system. 

Drawings and 
Diagrams 

65666,65673 

Other 

 

 
 

Summary of Contents for EX378B02

Page 1: ...ystem Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mai...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...in which the instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument St...

Page 7: ...owered with a 24V or a 12V phantom power supply but this will limit the maximum output voltage Externally Polarized Microphone System Prepolarized Microphone System Phantom Powered Microphone System T...

Page 8: ...em such as the preamplifier signal conditioner or instrumentation Conversely the Cartridge Thermal Noise CTN specification provides the lowest measurable sound pressure level that can be detected abov...

Page 9: ...rugged grid cap shown below and as an independent part of a rugged kit to extend the lifespan of the precision measuring device caution should be taken to ensure the delicate diaphragm sensing element...

Page 10: ...ustry standards ensures accurate calibra tion against relevant specifications We also carry a line of acoustic calibrators that can be used to calibrate microphone sensitivity on site as needed Hand c...

Page 11: ...ow 1 to 2 minutes for the system to stabilize The meter or LED on the signal conditioner should be reading green This indicates proper operation and you may begin taking measurements If a faulty condi...

Page 12: ...outine calibration of sensors by the manufacturer is recommended as this helps build confidence in measurement accuracy and acquired data Danger Areas for pressure relief devices N A not a pressure re...

Page 13: ...14043 3425 Walden Ave Depew New York 14043 No 65238 No 3 of 3 Rev NR ECO 45769 Note Literature such as the manual or marketing materials describing the equipment or protective system must not contrad...

Page 14: ...d a voltage be supplied to an ICP Microphone System without a current regulating diode or equivalent electrical circuit This may include ohmmeters multi meters and continuity testers Safe Use After co...

Page 15: ...highly recommended to keep spare cables on hand to enable continuation of the test in the event of a cable failure Install Microphone system per drawing 68216 Adjustment The EX378XYYY series come ass...

Page 16: ...IV 1 GROUPS A B C D CODE IDENT NO 52681 DWG NO SCALE SHEET DRAWN CHECKED ENGINEER TITLE UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES DECIMALS X 05 XX 01 XXX 005 ANGLES 2 DEGREES FILLETS AND RAD...

Page 17: ...oltage 7 Vpk 7 Vpk Output Impedance 50 Ohm 50 Ohm Physical Housing Material Stainless Alloy Stainless Alloy Venting Rear Rear 4 Electrical Connector BNC Jack BNC Jack Mounting Thread Grid 0 5 60 UNS 0...

Page 18: ...MICROPHONE DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 3 1 18 KRM 3 1 18 GGS 3 1 18 66401 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any r...

Page 19: ...ia IIC T4 Ga I M1 Ex ia I CODE IDENT NO 52681 DWG NO SCALE SHEET DRAWN CHECKED ENGINEER TITLE UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES DECIMALS X 05 XX 01 XXX 005 ANGLES 2 DEGREES FILLETS...

Page 20: ...stem Type Ex378XYYY Optional accessory Type of Protection Ex ia Marking Ex ia I Ma Ex ia IIC T4 Ga Approved for issue on behalf of the lECEx Certification Body Position Signature for printed version D...

Page 21: ...ns as set out in lECEx Scheme Rules lECEx 02 and Operational Documents as amended STANDARDS The electrical apparatus and any acceptable variations to it specified in the schedule of this certificate a...

Page 22: ...e the interface to an external cable assembly The equipment is intended to be used for leak detection and acoustic noise tests in gaseous or mining explosive areas SPECIFIC CONDITIONS OF USE YES as sh...

Page 23: ...trinsic safety 1 EN 61326 1 2013 Electrical Equipment for Measurement Control and Laboratory Use EMC EN 61326 2 3 2013 Electrical Equipment for Measurement Control and Laboratory Use EMC EN 61010 1 20...

Page 24: ...venue du G n ral Leclerc FR 92260 Fontenay aux Roses Tel 33 1 40 95 60 60 Fax 33 1 40 95 86 56 I the undersigned hereby declare that the equipment specified above conforms to the a W Direce Standard s...

Page 25: ...nfidentiel s No report s N 144457 692709 8 Le respect des Exigences Essentielles de S curit et de Compliance with the Essential Health and Safety Sante est assure par la conformit a Requirements has b...

Page 26: ...nd a preamplifier The delivered signal from lOP Microphone Systems provides a typical 1 7V peak voltage output signal when making acoustic measurements The microphone cartridge is the sensing element...

Page 27: ...r metallic body and cable shield are reliably connected to the system earth The apparatus must be connected according to instruction manual For final installation the user shall take all necessary pre...

Reviews: