MODELS 483C30, 483C50, 482M179 AND 483M217 GENERAL OPERATION MANUAL
11
•
Now you can use the assigned IP address to address the unit.
Important Note: The communication protocol requires a unit id as part of the command header. The unit id is
not the IP address. To send commands to the unit Via Ethernet you must address the TCP-IP packets with the
proper IP address and ensure the packet payload contains the correct Unit Id in the command header.
The XPort device should be configured properly for communication with the signal conditioner. If problems
occur please check the communication parameters of the device as shown below.
Navigate to the actual device as shown in the left panel above and then click on the ‘Web Configuration’ tab and clock
on the ‘Go’ button. This will cause the following screen to appear, just select ‘OK’. There are no login parameters
defined.