background image

3.0 INSTALLATION 
 
Connect the 422E5X to the transducer 
with low noise cable (003 type only). 
Standard coaxial cable may be used 
between the amplifier and power unit, 
and between the power unit and readout 
device. (Fig 2) 
 

Fig 2 

 
CAUTION:  
Excessive accumulated charges on the 
input cable can destroy the field effect 
input transistor. These charges should be 
grounded out by shorting the center pin 
of the connector plug to the outer cable 
conductor with a metallic object before 
connecting them to the 422E5X charge 
amp. Care should be taken to avoid 
touching the input terminal to any object 
that could possess an electric charge 
(such as your fingers). 
 
 
4.0 OPERATION 
 
After connecting the 422E5X as 
indicated in Fig 2 Energize the system 
by turning on the power unit and waiting 
a few minutes for the amplifier to turn 
on.  
 
 
 
 

For a simple, self-test circuit, monitor 
the +9 to +12 V DC bias existing at the 
output during normal operation with a 
high impedance voltmeter. Many PCB 
power units feature such a test circuit 
(Fig 3). 
 

 

Fig 3 

 A typical bias meter. 

 
4.1 SYSTEM SENSITIVITY 
 
When the sensitivity (pC/unit) of the 
transducer is known (via calibration) 
along with the transfer sensitivity of the 
charge amp (mV/pC), the system 
sensitivity of the system can be 
calculated in the following manner: 
 
System sens = (Xdcr Sens)(Amp Sens) 
 
System sens is given in (mV/Unit) 
Xdcr Sens is given in (pC/Unit) 
Amp Sens is given in (mV/pC) 
 
 
5.0 MAINTENANCE AND REPAIR 
 
The 422E5X charge amplifier is 
hermetically sealed, and requires no 
maintenance. If the connectors become 
dirty, clean with lint-free wipes 
dampened with isopropyl alcohol. 
 
For extremely dirty or humid 
environments, the cable connections can 
be protected with low temperature heat 
shrink tubing. 
 
Should trouble occur, the factory should 
be contacted for assistance. 

MANUAL NUMBER: 25578 
MANUAL REVISION:  NR 

 

Reviews: